60 Watts
xppower.com
PDM60 Series
•
•
•
•
•
•
•
Medical & Industrial/IT Approvals
4000 VAC Isolation
Class II Construction
Single Outputs from 12 V to 48 V
High Efficiency
Non-standard Connectors Available
Operation to +60 °C
Specification
Input
General
Input Voltage
Input Frequency
Input Current
Inrush Current
• 85-264 VAC
Efficiency
• 80ꢀ typical
• 47-63 Hz
Isolation
• 4000 VAC Input to Output
• 55 kHz typical
• 2.7 W/in3
• 1.5 A rms max at 85 VAC
Switching Frequency
Power Density
MTBF
• 30/60 A max at 115/230 VAC
cold start at +25 ºC
• >420 kHrs per MIL-HDBK-217F,
+40 ºC ground benign
Power Factor
• Compliant with EN61000-3-2 Class A
equipment
Environmental
Input Protection
• Internal fuse in live line
Operating Temperature • -10 ºC to +60 ºC, derate above +40 ºC by
Output
2.5ꢀ/ºC to 50ꢀ load at +60 ºC
Output Voltage
Initial Set Accuracy
Minimum Load
Start Up Delay
Start Up Rise Time
Hold Up Time
• See table
Operating Humidity
Storage Temperature
Operating Altitude
Shock
• 5-95ꢀ RH, non-condensing
• -20 ºC to +85 ºC
•
1ꢀ at 50ꢀ load
• No minimum load required
• <500 ms
• 3000 m max
• 30 g pk, half sine, 6 axes
• 2 g, 5-500 Hz, 3 axes
• 40 ms typical
Vibration
• 12 ms min at 115 VAC & 100ꢀ load
• < 0.2ꢀ after 20 minute warm up
EMC & Safety
Drift
Emissions
• EN55022, level B conducted
EN55022, level B radiated,
EN60601-1-2, EN61204-3,
FCC 20780 level B
Line Regulation
Load Regulation
Transient Response
•
•
0.5ꢀ
3ꢀ
• 4ꢀ max. deviation, recovery to within
1ꢀ in 500 µs for a 25ꢀ load change
at 1 A/µs
Harmonic Currents
Voltage Flicker
• EN61000-3-2, class A
• EN61000-3-3
Ripple & Noise
• 1ꢀ pk-pk, 20 MHz BW (see note 2)
ESD Immunity
• EN61000-4-2, level 3 Perf Criteria A
Overvoltage Protection • 115-150ꢀ, recycle input to reset
Overload Protection • 120-170ꢀ, auto recovery
Short Circuit Protection • Trip and restart (Hiccup mode) indefinite
Radiated Immunity
• EN61000-4-3, 10 V/m, 80ꢀ mod
Perf Criteria A
EFT/Burst
• EN61000-4-4, level 3 Perf Criteria A
• EN61000-4-5, level 3 Perf Criteria A
Temperature
Coefficient
•
0.05ꢀ/ºC
Surge
Conducted Immunity
• EN61000-4-6, 10 Vrms, 80ꢀ mod
Perf Criteria A
T
Dips & Interruptions
Safety Approvals
• EN61000-4-11, 70ꢀ U 10 ms,
T
T
40ꢀ U 100 ms, <5ꢀ U 5000 ms
Perf Criteria A, B, B
• EN60601-1-1, UL60601-1-1:2003,
CSA60601-1, EN60950-1:2001,
UL/cUL60950, CE Mark, CCC Mark