AHP1501 series
15W Medical Power Adapters with Interchangeable Input Plugs
INPUTꢀSPECIFICATIONSꢀ
ENVIRONMENTALꢀSPECIFICATIONSꢀ
Input Voltage Range
80-275 VAC (100-240V Nominal)
Oper. Temperature
-10 to +40°C (FL) *
50% Load @ 70°C
See Output Derate Curve, Fig. 2
-40 to +85°C *
See Input Derate curve, Fig. 1
Frequency Range (Sine Wave) 47-63 Hz
Input Current Low Line (100VAC)
High Line (240VAC)
0.6A max.
0.4A max.
35A max.
85A max.
Storage Temperature
Operating/Storage Humidity
Cooling
0% to 95% RH *
Inrush Current, typ (Cold Start) Low Line (100VAC)
High Line (240VAC)
Free Air Convection
EN55011 (CISPR11),
EN61000-3-2,-3 Class B
2kV, All Conditions
8kV, IEC61000-4-2
6kV, IEC61000-4-2
100K h
EMC
Protection Class
Double Insulated, Class II
Input NL Power Consumption
See Selection Chart
Surge Voltage
ESD Air
ESD Contact
OUTPUTꢀSPECIFICATIONSꢀ
MTBF MIL-HDBK-217F, 25°C
Operating Altitude
Vibration
3000m
Voltage and Current
See Selection Chart
0 to FL Rating
5G, 10~500Hz, 10 min./1 cycle,
60 min. each along X, Y, Z axes
UL94V-1
Output Power Range
Total Regulation (Note 2)
Line Regulation, FL (Note 1)
Over Load Protection*
Short Circuit Protection
Transient Response
See Selection Chart
± 1%, max.
Flammability Rating
110-150%, auto recover
Auto Recover
NOTESꢀ
110Vin, FL
4mS
Temperature Coefficient
Ripple/Noise (Note 3)
Start Up / Hold Up Time
±0.04%/°C, typ
See Selection Chart
1. Line Regulation defined as changing the input voltage ±10% from
nominal line at rated load.
100Vin, FL
2S max. / 12mS min.
2. Total Regulation is defined by as changing the load ±40% from 60%
rated load.
3. Ripple & Noise is measured by using a 20MHz bandwidth limited
oscilloscope and terminating the output with a 0.47uF capacitor at
rated load and nominal line.
GENERALꢀSPECIFICATIONSꢀ
Input-Out Isolation
Efficiency(FL, Nom. Input))
Safety
4000 VAC
See Selection Chart
ANSI/AAMI ES 60601-1:2005
(UL/cUL 3rd Edition)
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
PHYSICALꢀSPECIFICATIONSꢀ
Size
2.36” x 1.71” x 1.58”
(60 x 43.4 x 40.1mm)
Wallmount Style
Construction
Weight
5.82oz (165g)
All specifications are typical at nominal input, full load, and 25°C unless other-
wise noted
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
WWW.ASTRODYNETDI.COM