5秒后页面跳转
AD595CQ PDF预览

AD595CQ

更新时间: 2024-01-20 22:16:33
品牌 Logo 应用领域
亚德诺 - ADI 放大器
页数 文件大小 规格书
8页 140K
描述
Monolithic Thermocouple Amplifiers with Cold Junction Compensation

AD595CQ 数据手册

 浏览型号AD595CQ的Datasheet PDF文件第2页浏览型号AD595CQ的Datasheet PDF文件第3页浏览型号AD595CQ的Datasheet PDF文件第4页浏览型号AD595CQ的Datasheet PDF文件第6页浏览型号AD595CQ的Datasheet PDF文件第7页浏览型号AD595CQ的Datasheet PDF文件第8页 
AD594/AD595  
The AD594/AD595 also includes an input open circuit detector  
that switches on an alarm transistor. This transistor is actually a  
current-limited output buffer, but can be used up to the limit as  
a switch transistor for either pull-up or pull-down operation of  
external alarms.  
this terminal can be produced with a resistor between –C and  
–T to balance an increase in +T, or a resistor from –T to COM  
to offset a decrease in +T.  
If the compensation is adjusted substantially to accommodate a  
different thermocouple type, its effect on the final output volt-  
age will increase or decrease in proportion. To restore the  
nominal output to 10 mV/°C the gain may be adjusted to match  
the new compensation and thermocouple input characteristics.  
When reducing the compensation the resistance between –T  
and COM automatically increases the gain to within 0.5% of the  
correct value. If a smaller gain is required, however, the nominal  
47 kinternal feedback resistor can be paralleled or replaced  
with an external resistor.  
The ice point compensation network has voltages available with  
positive and negative temperature coefficients. These voltages  
may be used with external resistors to modify the ice point com-  
pensation and recalibrate the AD594/AD595 as described in the  
next column.  
The feedback resistor is separately pinned out so that its value  
can be padded with a series resistor, or replaced with an external  
resistor between Pins 5 and 9. External availability of the feedback  
resistor allows gain to be adjusted, and also permits the AD594/  
AD595 to operate in a switching mode for setpoint operation.  
Fine calibration adjustments will require temperature response  
measurements of individual devices to assure accuracy. Major  
reconfigurations for other thermocouple types can be achieved  
without seriously compromising initial calibration accuracy, so  
long as the procedure is done at a fixed temperature using the  
factory calibration as a reference. It should be noted that inter-  
mediate recalibration conditions may require the use of a  
negative supply.  
CAUTIONS:  
The temperature compensation terminals (+C and –C) at Pins 2  
and 6 are provided to supply small calibration currents only. The  
AD594/AD595 may be permanently damaged if they are  
grounded or connected to a low impedance.  
The AD594/AD595 is internally frequency compensated for feed-  
back ratios (corresponding to normal signal gain) of 75 or more.  
If a lower gain is desired, additional frequency compensation  
should be added in the form of a 300 pF capacitor from Pin 10  
to the output at Pin 9. As shown in Figure 5 an additional 0.01 µF  
capacitor between Pins 10 and 11 is recommended.  
EXAMPLE: TYPE E RECALIBRATION—AD594/AD595  
Both the AD594 and AD595 can be configured to condition the  
output of a type E (chromel-constantan) thermocouple. Tem-  
perature characteristics of type E thermocouples differ less from  
type J, than from type K, therefore the AD594 is preferred for  
recalibration.  
While maintaining the device at a constant temperature follow  
the recalibration steps given here. First, measure the device  
temperature by tying both inputs to common (or a selected  
common-mode potential) and connecting FB to VO. The AD594  
is now in the stand alone Celsius thermometer mode. For this  
example assume the ambient is 24°C and the initial output VO  
is 240 mV. Check the output at VO to verify that it corresponds  
to the temperature of the device.  
AD594/  
AD595  
9
VO  
300pF  
COMP  
10  
11  
0.01F  
+V  
Figure 5. Low Gain Frequency Compensation  
RECALIBRATION PRINCIPLES AND LIMITATIONS  
The ice point compensation network of the AD594/AD595  
produces a differential signal which is zero at 0°C and corre-  
sponds to the output of an ice referenced thermocouple at the  
temperature of the chip. The positive TC output of the circuit is  
proportional to Kelvin temperature and appears as a voltage at  
+T. It is possible to decrease this signal by loading it with a  
resistor from +T to COM, or increase it with a pull-up resistor  
from +T to the larger positive TC voltage at +C. Note that  
adjustments to +T should be made by measuring the voltage which  
tracks it at –T. To avoid destabilizing the feedback amplifier the  
measuring instrument should be isolated by a few thousand  
ohms in series with the lead connected to –T.  
Next, measure the voltage –T at Pin 5 with a high impedance  
DVM (capacitance should be isolated by a few thousand ohms  
of resistance at the measured terminals). At 24°C the –T voltage  
will be about 8.3 mV. To adjust the compensation of an AD594  
to a type E thermocouple a resistor, R1, should be connected  
between +T and +C, Pins 2 and 3, to raise the voltage at –T by  
the ratio of thermocouple sensitivities. The ratio for converting a  
type J device to a type E characteristic is:  
r (AD594) =(60.9 µV/°C)/(51.7 µV/°C)= 1.18  
Thus, multiply the initial voltage measured at –T by r and ex-  
perimentally determine the R1 value required to raise –T to that  
level. For the example the new –T voltage should be about 9.8 mV.  
The resistance value should be approximately 1.8 k.  
+IN  
1
+T  
3
4
5
The zero differential point must now be shifted back to 0°C.  
This is accomplished by multiplying the original output voltage  
VO by r and adjusting the measured output voltage to this value  
by experimentally adding a resistor, R2, between –C and –T,  
Pins 5 and 6. The target output value in this case should be  
about 283 mV. The resistance value of R2 should be approxi-  
mately 240 k.  
AD594/  
–IN AD595  
14  
8
COM  
–T  
FB  
VO  
9
Figure 6. Decreased Sensitivity Adjustment  
Changing the positive TC half of the differential output of the  
compensation scheme shifts the zero point away from 0°C. The  
zero can be restored by adjusting the current flow into the nega-  
tive input of the feedback amplifier, the –T pin. A current into  
Finally, the gain must be recalibrated such that the output VO  
indicates the device’s temperature once again. Do this by adding  
a third resistor, R3, between FB and –T, Pins 8 and 5. VO should  
now be back to the initial 240 mV reading. The resistance value  
REV. C  
–5–  

与AD595CQ相关器件

型号 品牌 描述 获取价格 数据表
AD596 ADI Thermocouple Conditioner and Setpoint Controller

获取价格

AD596* ETC Thermocouple Conditioner and Setpoint Controller

获取价格

AD-5960S BOTHHAND ADSL LINE TRANSFORMER

获取价格

AD-5961S BOTHHAND ADSL LINE TRANSFORMER

获取价格

AD5962-9313001MPA ADI IC 1 CHANNEL, VIDEO AMPLIFIER, CDIP8, Audio/Video Amplifier

获取价格

AD596AH ADI Thermocouple Conditioner and Setpoint Controller

获取价格