5秒后页面跳转
AD595AQ PDF预览

AD595AQ

更新时间: 2024-01-28 20:33:37
品牌 Logo 应用领域
亚德诺 - ADI 放大器
页数 文件大小 规格书
8页 140K
描述
Monolithic Thermocouple Amplifiers with Cold Junction Compensation

AD595AQ 数据手册

 浏览型号AD595AQ的Datasheet PDF文件第2页浏览型号AD595AQ的Datasheet PDF文件第3页浏览型号AD595AQ的Datasheet PDF文件第4页浏览型号AD595AQ的Datasheet PDF文件第5页浏览型号AD595AQ的Datasheet PDF文件第7页浏览型号AD595AQ的Datasheet PDF文件第8页 
AD594/AD595  
of R3 should be approximately 280 k. The final connection  
diagram is shown in Figure 7. An approximate verification of  
the effectiveness of recalibration is to measure the differential  
gain to the output. For type E it should be 164.2.  
THERMAL ENVIRONMENT EFFECTS  
The inherent low power dissipation of the AD594/AD595 and  
the low thermal resistance of the package make self-heating  
errors almost negligible. For example, in still air the chip to am-  
bient thermal resistance is about 80°C/watt (for the D package).  
At the nominal dissipation of 800 µW the self-heating in free air  
is less than 0.065°C. Submerged in fluorinert liquid (unstirred)  
the thermal resistance is about 40°C/watt, resulting in a self-  
heating error of about 0.032°C.  
+IN  
COM  
+T  
1
4
3
14  
–IN  
R1  
R2  
AD594/  
AD595  
+C  
–C  
2
6
SETPOINT CONTROLLER  
The AD594/AD595 can readily be connected as a setpoint  
controller as shown in Figure 9.  
9
8
VO  
FB  
–T  
5
LOW = > T < SETPOINT  
HIGH = > T > SETPOINT  
HEATER  
DRIVER  
TEMPERATURE  
R3  
COMPARATOR OUT  
CONSTANTAN  
SETPOINT  
VOLTAGE  
HEATER  
+5V  
(ALUMEL)  
Figure 7. Type E Recalibration  
INPUT  
13  
12  
11  
10  
9
8
14  
When implementing a similar recalibration procedure for the  
AD595 the values for R1, R2, R3 and r will be approximately  
650 , 84 k, 93 kand 1.51, respectively. Power consump-  
tion will increase by about 50% when using the AD595 with  
type E inputs.  
OVERLOAD  
DETECT  
20M⍀  
(OPTIONAL)  
FOR  
AD594/  
AD595  
+A  
HYSTERESIS  
ICE  
POINT  
COMP.  
G
G
–TC  
+TC  
Note that during this procedure it is crucial to maintain the  
AD594/AD595 at a stable temperature because it is used as the  
temperature reference. Contact with fingers or any tools not at  
ambient temperature will quickly produce errors. Radiational  
heating from a change in lighting or approach of a soldering iron  
must also be guarded against.  
IRON  
1
2
3
4
5
6
7
(CHROMEL)  
TEMPERATURE  
CONTROLLED  
REGION  
COMMON  
Figure 9. Setpoint Controller  
The thermocouple is used to sense the unknown temperature  
and provide a thermal EMF to the input of the AD594/AD595.  
The signal is cold junction compensated, amplified to 10 mV/°C  
and compared to an external setpoint voltage applied by the  
user to the feedback at Pin 8. Table I lists the correspondence  
between setpoint voltage and temperature, accounting for the  
nonlinearity of the measurement thermocouple. If the setpoint  
temperature range is within the operating range (–55°C to  
+125°C) of the AD594/AD595, the chip can be used as the  
transducer for the circuit by shorting the inputs together and  
utilizing the nominal calibration of 10 mV/°C. This is the centi-  
grade thermometer configuration as shown in Figure 13.  
USING TYPE T THERMOCOUPLES WITH THE AD595  
Because of the similarity of thermal EMFs in the 0°C to +50°C  
range between type K and type T thermocouples, the AD595  
can be directly used with both types of inputs. Within this ambi-  
ent temperature range the AD595 should exhibit no more than  
an additional 0.2°C output calibration error when used with  
type T inputs. The error arises because the ice point compensa-  
tor is trimmed to type K characteristics at 25°C. To calculate  
the AD595 output values over the recommended –200°C to  
+350°C range for type T thermocouples, simply use the ANSI  
thermocouple voltages referred to 0°C and the output equation  
given on page 2 for the AD595. Because of the relatively large  
nonlinearities associated with type T thermocouples the output  
will deviate widely from the nominal 10 mV/°C. However, cold  
junction compensation over the rated 0°C to +50°C ambient  
will remain accurate.  
In operation if the setpoint voltage is above the voltage corre-  
sponding to the temperature being measured the output swings  
low to approximately zero volts. Conversely, when the tempera-  
ture rises above the setpoint voltage the output switches to  
the positive limit of about 4 volts with a +5 V supply. Figure  
9 shows the setpoint comparator configuration complete with a  
heater element driver circuit being controlled by the AD594/  
AD595 toggled output. Hysteresis can be introduced by inject-  
ing a current into the positive input of the feedback amplifier  
when the output is toggled high. With an AD594 about 200 nA  
into the +T terminal provides 1°C of hysteresis. When using a  
single 5 V supply with an AD594, a 20 Mresistor from VO to  
+T will supply the 200 nA of current when the output is forced  
high (about 4 V). To widen the hysteresis band decrease the  
resistance connected from VO to +T.  
STABILITY OVER TEMPERATURE  
Each AD594/AD595 is tested for error over temperature with  
the measuring thermocouple at 0°C. The combined effects of  
cold junction compensation error, amplifier offset drift and gain  
error determine the stability of the AD594/AD595 output over  
the rated ambient temperature range. Figure 8 shows an AD594/  
AD595 drift error envelope. The slope of this figure has units  
of °C/°C.  
+0.6؇C  
0
50؇C  
25؇C  
–0.6؇C  
TEMPERATURE OF AD594C/AD595C  
Figure 8. Drift Error vs. Temperature  
–6–  
REV. C  

与AD595AQ相关器件

型号 品牌 描述 获取价格 数据表
AD595C ADI Monolithic Thermocouple Amplifiers with Cold Junction Compensation

获取价格

AD595CD ADI Monolithic Thermocouple Amplifiers with Cold Junction Compensation

获取价格

AD595CDZ ADI Monolithic Thermocouple Amplifiers

获取价格

AD595CQ ADI Monolithic Thermocouple Amplifiers with Cold Junction Compensation

获取价格

AD596 ADI Thermocouple Conditioner and Setpoint Controller

获取价格

AD596* ETC Thermocouple Conditioner and Setpoint Controller

获取价格