2.7 GHz RF Signal Analyzer
Digitizer Front Panel
IF input (normally connected to PXI-5600 IF output), AC coupled
Impedance/connector..................... 50 Ω, SMA female
Input power..................................... 0 dBm nominal, +10 dBm
full-scale
Maximum safe input level.............. +20 dBm
Maximum DC input voltage............ 2 V
VSWR.............................................. <1.5:1, 5 to 25 MHz
External frequency reference input (normally connected to
PXI-5600 10 MHz output)
Impedance/connector..................... 50 Ω, SMA female
Input power..................................... -5 to +15 dBm
Maximum safe input level.............. +16 dBm
Maximum DC input voltage............ 10 V
Input frequency range .................... 10 MHz 0.5 ppm
External trigger (PFI 1)
Environment
Warm-up time..................................... 20 minutes
Operating temperature ....................... 0 to 50 °C
Relative humidity............................ 10 to 90%, noncondensing
Storage temperature........................... -20 to +70 °C
Relative humidity............................ 5 to 95%, noncondensing
Safety and Compliance
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1, CSA 61010-1
Note: For UL and other safety certifications, refer to the product label or
visit ni.com/certification, search by model number or product line, and
click the appropriate link in the Certification column.
Connector........................................ SMB male
Level................................................ TTL
Maximum input voltage ................. 5.5 V
Electromagnetic Compatibility
This product is designed to meet the requirements of the following
standards of EMC for electrical equipment for measurement, control,
and laboratory use:
IF/Baseband
Resolution ........................................... 14 bits
IF input level........................................ 0 dBm nominal, +10 dBm
full-scale
IF frequency range .............................. 5 to 25 MHz
Sample rate......................................... 64 MS/s and integer divisions
down to 1 kS/s
• EN 61326 EMC requirements; Minimum Immunity
• EN 55011 Emissions; Group 1, Class A
• CE, C-Tick, ICES, and FCC Part 15 Emissions; Class A
Onboard memory................................. 16/32 MS option
Using DDC (complex data).................. 8/16 MS option
Note: For EMC compliance, operate this device according to
product documentation.
Power Requirements (Typical)
CE Compliance
This product meets the essential requirements of applicable European
Directives, as amended for CE marking, as follows:
+3.3 VDC ( 5ꢀ) +5 VDC ( 5ꢀ) +12 VDC ( 5ꢀ) -12 VDC ( 5ꢀ)
Downconverter (PXI-5600)
Digitizer (PXI-5620)
920 mA
600 mA
2.3 A
1.5 A
700 mA
450 mA
115 mA
35 mA
• 2006/95/EC; Low-Voltage Directive (safety)
• 2004/108/EC; Electromagnetic Compatibility Directive (EMC)
Calibration
Interval ................................................ 1 year
Note: Refer to the Declaration of Conformity (DoC) for this product for
any additional regulatory compliance information. To obtain the DoC for
this product, visit ni.com/certification, search by model number or
product line, and click the appropriate link in the Certification column.
Physical
PXI-5600 ............................................. 3U, 3-slot, PXI/Compact PCI
module 21.6 by 6.0 by 13.0 cm
(8.5 by 2.4 by 5.1 in.)
Environmental Management
PXI-5620 ............................................. 3U, 1-slot, PXI/Compact PCI
module 21.6 by 2.0 by 13.0 cm
NI is committed to designing and manufacturing products in an
environmentally responsible manner. NI recognizes that eliminating
certain hazardous substances from our products is beneficial not only
to the environment but also to NI customers.
(8.5 by 0.8 by 5.1 in.)
For additional environmental information, refer to the NI and the
Environment Web page at ni.com/environment. This page contains
the environmental regulations and directives with which NI complies, as
well as other environmental information not included in this document.
BUY ONLINE at ni.com or CALL 800 813 3693 (U.S.)