SDA-4040D STEALTH DIGITAL ANALYZER
7
continued
Specifications
Measurable QAM ingress
64 QAM
View option (OPT 4)
Zero Span mode
–25 to –40 dBc
–30 to –40 dBc
3.0 dB
The QAM View option can be factory installed in any new or
existing SDA Series instrument.The specifications and features
are in addition to the standard measurement features of the
SDA Series.When ordering, please specify OPT 4A for 8MHz,
DVB-C, ITU-T J.83 Annex A, or OPT 4B for 6MHz, DVS-031, ITU-T
J.83 Annex B.
Video BW
Resolution BW
>1 MHz, 100 kHz, 10 kHz, 100 Hz
2 MHz, 280 kHz, 30 kHz
256 QAM
Accuracy
Graphic display
Measurement BW Compensation
programmable 1 kHz to 99 MHz
Pulse measurement accuracy
Digital summary (including MER/EVM, Pre/Post FEC BER,
Equalizer Stress, Carrier Offset, Symbol Rate) with limit/margin
test results, QAM level. IQ constellation with zoom. Adaptive
Equalizer Display (8 feed forward/24 feedback), Frequency
Response, Group Delay. Ingress/Noise Under the Carrier.
Power source
Note: Option powered from SDA Series nickel metal hydride
battery. Operating time is specified for continuous use in QAM
View mode. Option includes high output charger.
nominal level in 10 µs
2 dB from nominal in 5 µs
(>1 MHz VBW, 280 kHz RBW)
100 µs to 20 s (1, 2, 5 settings)
Modulation type
64/256 QAM, DVB-C,
ITU-T J.83 Annex A (OPT 4A)
64/256 QAM, DVS-031,
Sweep times
Intermodulation distortion (CSO/CTB)
ITU-T J.83 Annex B (OPT 4B)
8 MHz (OPT 4A); 6 MHz (OPT 4B)
Range(5)
Resolution
≥ 60 dB
0.1 dB
Channel bandwidth
Measurable input range (lock range)
64 QAM
–20 to +50 dBmV (typical)
Charge time
~4 hours
2.5 hours continuous use (typical)
Data storage
256 QAM
–15 to +50 dBmV (typical)
Operating time
Universal AC charger/adapter
Input
Output
Physical dimensions
Frequency tuning
Files types that can be stored: Sweepless Sweep, autotests, tilt
graphs, channel plans, and scan graphs. Spectrum graphs
(normal, normal with max hold, and CSO/CTB) can also be
stored. Memory space is allocated on demand. All files stored
as data, not as screen picture.Typical mix of files for 78-
channel plan, 8 channel plans; 16 sweep references, 80 sweep
traces; 40 scan files, 20 spectrum displays; 20 autotests
50 to 860 MHz (Digital QAM mode)
Resolution
100-250 VAC, 50-60 Hz, 0.5 A
Charge 15 V at 750 ma
50 kHz
BER (bit error rate)
64 QAM Pre-FEC/OPTs 4A and 4B
64 QAM Post-FEC/OPTs 4A and 4B
256 QAM Pre-FEC/OPT 4A and 4B
256 QAM Post-FEC/OPT 4A and 4B
MER (modulation error ratio)
64 QAM/Option 4A
10–4 to 10–9
10–4 to 10–9
10–4 to 10–9
10–4 to 10–9
(total SDA-5000 size with OPT 4)
6 x 10.5 x 4.25 in (15.2 x 26.7 x 10.8 cm)
Weight
Approx. 7.7 lb (3.5 kg)
–4 to 113°F (–20 to 45°C)
Operating temperature range
Serial interface
22 to 35 dB
Notes
RS232; Epson, IBM, Seiko, and Diconix printers
(1)
Typical specifications
Accuracy
2.0 dB (typical, see chart below)
(2)
Relative to 25°C
Input configuration
64 QAM/Option 4B
Accuracy
256 QAM/Option 4A
Accuracy
256 QAM/Option 4B
Accuracy
21 to 35 dB
1.5 dB
28 to 35 dB
(3)
At 25°C and +20 dBmV
(4)
Connector type
75 ΩType F Female
(Optional 75 ΩType BNC Female)
100 VAC, 140 VDC
60
Rango de límites de medida
50
40
30
2.0 dB (typical, see chart below)
28 to 35 dB
Maximum sustained voltage
2.0 dB Accuracy
PathTrak field view (OPT 3 required)
1.5 dB
EVM (error vector magnitude)
64 QAM/Option 4A
Accuracy
Update rate
2x/second (remote trace)
~1x/second (local trace)
5/1/2/5/10/20dB/div.
20
10
0
1.2% to 5.2%
0.5% (1.2% to 2.0%)
1.0% (2.1% to 4.0%)
1.4% (4.1% to 5.2%)
1.2% to 5.8%
0.5% (1.2% to 2.5%)
1.1% (2.6% to 5.8%)
1.1% to 2.5%
Display scaling
Selectable nodes
14 (selectable via PathTrak HCU)
24
–10
0
10
14
Nivel portadora de video (dBmV)
(5)
64 QAM/Option 4B
Accuracy
60
50
Rango de límites de medida
256 QAM/Option 4A
Accuracy
256 QAM/Option 4B
Accuracy
40
30
2.0 dB Accuracy
0.6%
1.1% to 2.5%
20
0.5%
QAM level measurement
Signal types
Range
10
0
64 QAM, 256 QAM
–20 to +45 dBmV
1.0 dB
–20
–10
0
10
20
30
Accuracy
Nivel portadora de video (dBmV)
Flatness
0.5 dB
Linearity
1.0 dB
Temperature
0.5 dB (typical)