Technical Data Sheet
Calibration Kit
RPC-1.85
08CK020-150
LRL Version
Documentation
This kit is delivered with
•
USB-Stick
Standard Definitions as data files for Vector Network Anaylzer Families PNA (Keysight/Agilent) and ZVA
(Rohde&Schwarz). Calibration Certificate as PDF-file.
Standard Definitions Cards
Printed Standard Definitions that can be used on nearly all Vector Network Analyzers.
Kit Info Card
Handling precautions and information for installing Standard Definitions on a Vector Network Analyzer.
Calibration Certificate
Details see “Declaration of calibration options”
•
•
•
•
Operating Manual
Electrical specifications
This specification covers electrical key values for the main items of the calibration kit. Specific datasheets
are available for each component among the part number.
Device
Frequency
Parameter
Specification
≤ 4 GHz
to
≤ 2.0°
≤ 5.0°
≤ 8.0°
DC
Shortsb
(plug and jack)
Error from
Nominal Phase
> 4 GHz to
> 40 GHz to
≤ 40 GHz
≤ 70 GHz
≤ 4 GHz
≤ 40 GHz
≤ 70 GHz
≥ 34 dB
≥ 26 dB
≥ 24 dB
0.5 GHz to
> 4 GHz to
> 40 GHz to
Precision air linesc
Return Loss
b.
c.
The specifications for the shorts are given as allowed deviation from nominal model as defined in calibration certificate included
with your kit.
Calculated value without warranty. The return loss specification includes the connector interfaces. The minimum frequency of
the air lines is based on calculations of the impedance change due to skin depth. Refer to the calibration certificate included with
your kit for the exact dimensions of your precision air lines.
Rosenberger Hochfrequenztechnik GmbH & Co. KG
P.O.Box 1260 D-84526 Tittmoning Germany
www.rosenberger.de
Tel. : +49 8684 18-0
Fax : +49 8684 18-499
Email : info@rosenberger.de