POWER TRANSFORMER
CHASSIS MOUNT :
TOROIDAL MEDICAL SERIES
VPM18-8800
Description:
The toroidal construction inherently reduces stray fields, increases efficiency and minimizes size
compared to traditional EI transformers. The addition of a Flux Band further reduces the remaining
stray fields. The shield between Primary and Secondary improves safety, reduces common
mode signals and minimizes leakage current. Built with a Class F (155°) insulation system.
140°C self-resetting thermal switch is included in each primary.
A
Electrical Specifications (@25C)
1. Maximum Power: 160VA
RoHS
2011/65/EU
Photo for illustration only
2. Input Voltages: 100, 120, 220, 240VAC, 50/60Hz
3. Output Voltages: 9VAC @17.60A or 18VAC CT @ 8.8A
4. Voltage Regulation: 7.7% TYP from full load to no load
5. Temperature Rise: 50°C TYP
6. Hipot: 4000VAC, Primary to Secondary, Primary & Secondary to Shield & mounting surface
7. Efficiency: 92% TYP. @ full load
8. Earth Leakage: ≤10µA (See Fig. 1), Patient Leakage: ≤ 25µA (see Fig. 2)
Agency File:
UL: File E122529, UL 60601-1/(R) 2012 Medical Electrical Equipment – Part 1 with 2 MOPP
CE: ES 60601-1 (IEC 60601-1:2005, MOD)
cUL: C22.2 No. 60601-1:14, Medical Electrical Equipment – Part 1
CB Certified.
OD
Dimensions: Inches (mm)
*Add 0.188 (3) to the height for mounting hardware
O.D.
I.D.
HT.*
FLAT
WASHER
RUBBER
PAD
BOLT
Weight: 1.8Kg
4.4 (113) 1.6(40)
2.0(52)
MOUNTING
METAL
PLATE
Mounting:
Transformer is provided with one metal mounting plate, two rubber pads, M6 x 60mm bolt, nut,
spring and flat washer.
H
Connections:
Transformer is provided with 8” (203mm) long, 0.25” (6.35mm) stripped and tinned, stranded
UL 1015 lead wire. Primaries are 22AWG, Secondaries are 12AWG, and Shield is 20AWG.
The GRN/YEL shield lead is typically grounded. Do not lift transformer by leads!
SPRING
WASHER
RUBBER
PAD
NUT
Input Options:
100VAC: Input to Gray & Blue, jumper White & Brown, jumper Blue & Violet.
120VAC: Input to White & Blue, jumper White & Brown, jumper Blue & Violet.
220VAC: Input to Gray & Violet, jumper Blue & Brown
240VAC: Input to White and Violet, jumper Blue & Brown
Output Options:
WHT
GRY
BLK
9VAC: Output from Black & Red, jumper Black & Orange, jumper Red to Yellow
18VAC: Output from Black & Yellow, jumper Red & Orange
RED
ORG
BLU
BRN
PRI
SEC
Primary and secondary windings are designed to be connected in series or parallel. Windings
are not intended to be used independently.
YEL
VIO
RoHS Compliance: Meets the requirements of 2011/65/EU, known as the RoHS 2 initiative.
GRN/YEL
SCHEMATIC
* At printing, this document is considered “uncontrolled”. Contact Triad Magnetics’ website for current version
Web: www.TriadMagnetics.com
Phone 951-277-0757
Fax 951-277-2757
460 Harley Knox Blvd.
Perris, California 92571
Publish Date: July 7, 2016