Total Power International, Inc.
50W Desk Top Type
Medical Power Supplies
TPMPU51 SERIES
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single to Triple Output
Optional Output Connector (See page appendix)
Over Voltage and Over Load protection
Class I
ANSI/AAMI ES 60601-1: 2005(UL/cUL 3rd Edition)
EN 60601-1:2006 (TUV/T-mark 3rd Edition)
Input to Output : 2MOPP
3 year warranty
Electrical Characteristics:
Safety Approvals Input Voltage Range
100~240VAC
Vin
Application:
Operate Voltage Range
90~264VAC
47~63Hz
- Blood Pressure system
- Portable medical devices
- ECG machine
Input Frequency
fin
Output Power Range
Output Voltage Range
Output Current Range
Input Current (Low Line)
Input Current (High Line)
See rating chart
See rating chart
See rating chart
Po
Vo
Io
Io=Full load, Vin=100VAC
Io=Full load, Vin=240VAC
Io=Full load, 25°C, Cool start, Vin=115VAC
Io=Full load, 25°C, Cool start, Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
0.91~1.04A
0.43~0.53A
35A (max.)
70A (max.)
Iil
Iih
Low Line Inrush Current
High Line Inrush Current
Efficiency
Ir
Safety Approvals:
75~88%
Eff
Line Regulation
0.5~1%
REG-i
REG-o
OVP
RoHS
COMPLIANT
EN 60601-1
Load Regulation
IEC 60601-1
2002/95/EC
Vin=230VAC
3~7%
Over Voltage Protection
112~132%
110~150%
4mS (max.)
16mS (min.)
OCP Over Current Protection
Time of Transient Response
Hold-Up Time
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Ttr
Th
Ts
Start Up Time
0.3~2S
Vp-p Ripple & Noise(Peak to Peak)
1% (max.)
0.25mA (typ.)
Safety Ground Leakage Current
Ilk
Vin=240VAC/60Hz
All output
Temperature Coefficient
TC
0.04%/°C
Operating Altitude
Up to 3000m
See rating chart
6321VDC (min.)
OA
Pno
Vps
No-Load Power Consumption
Dielectric Withstanding Voltage
No load, Vin=230VAC
Primary to secondary
Dielectric Withstanding Voltage
for Primary to PE
Vpg
Ris
Primary to PE
2121VDC (min.)
Test Voltage=500VDC
(min.)
50M
Isolation Resistance
Note: The Ripple & Noise which is under 3.3VDC at 2% max.
Environmental
Operating Temperature
See derating curve
-40~85°C
To
Ts
Storage Temperature
Operating Humidity
0~95%
Ho
Storage Humidity
0~95%
Hr
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F
Derate linearly from 100% load at 50°C to 50% load at 70°C
0.1M Hrs (min.)
MTBF
Pd
Total Power International, Inc
Toll Free: 877-646-0900
www.total-power.com