Total Power International, Inc.
UNIVERSAL INPUT AC-DC MEDICAL APPLICATION OPEN FRAME
SINGLE OUTPUT 43 WATTS INTERNAL SWITCHING
POWER SUPPLIES TPMI43 SERIES
FEATURES:
ACCOMMODATE UNIVERSALAC SOURCES
COMPACT 4”x2”x1” PCB FORMAT
MEET MEDICAL STANDARD EN 60601-1/
UL 2601-1
EMI MEET EN 55011 & FCC CLASS B
CE MARKING COMPLIANCE
SPECIFICATION
INPUT SPECIFICATION
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Connector: V-M connector.
Output Wattage: Typ. 42 watts.
Input Voltage: Range typ. 90-264Vac.
Rating 100 to 240Vac.
Input Connector: V-M connector.
Line Regulation: Typ. 0.2%.
Input Frequency: 47-63Hz.
Load Regulation: Typ. ±2-3%.
Inrush Current: ≤12A rms cold start @230Vac.
Input Current: Typ. 0.7A@115Vac, 0.35A @230Vac.
Dielectric Withstand: Meet IEC 60601-1.
EMI: Meet EN55011 & FCC Class B.
Hold-up Time: Typ. 20mS @115Vac, 100mS @ 230Vac.
Earth Leakage: Less than 0.1mA @230Vac.
Noise & Ripple: Typ. 1.0% peak to peak.
OVP: Built-in at main output (Latch).
Adjustability: Factory set.
Overload Protection (OLP):
Fully protected against outputs overload and short circuit.
Typ. OLP set at 105-120% of rating output power.
Consult the factory for OLP setting.
GENERAL SPECIFICATION
Efficiency: Typ. 85-90%(Various with output voltage).
Switching Frequency: 85 KHz.
Operating Temperature: 0 to +50 °C under forced air flow
for full load. Derate by 1.3W/ °C
Circuit Topology: Fixed frequency flyback circuit.
Transient Response: Output voltage returns in less than
0.5mS following a 50% load change.
Safety Standard: Meet Medical standards UL 2601-1/
EN 60601-1 Class I.
from 40°C to 50 °C.
Storage Temperature: -20 to +85°C.
Temperature Coefficient: Typ. 0.04%/°C.
Cooling: At least 5 cfm(100LFM) direct forward air flow is
required to achieve full rating power.
Power Density: 5.5Watts / Cubic Inch.
Construction: 4”x2”x1” PCB format.
Note: Due to requests in market and advances in technology, specifications subject to change without notification.
For the details of safety approval, please consult the factory.
Total Power International, Inc Toll Free: 877-646-0900(US) PH# (978)453-7272 www.total-power.com