Total Power International, Inc.
30W Desk Top Type
Medical Power Supplies
TPHPU32A SERIES
Features:
Wide Operating Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C18 Input Inlet
Single Output
Over Voltage and Over Load protection
ANSI/AAMI ES 60601-1: 2005(UL/cUL 3rd Edition)
EN 60601-1:2006 (TUV/T-mark 3rd Edition)
Input to Output : 2MOPP
Energy Star 2.0, Efficiency level V
Class II
2 year warranty
Electrical Characteristics:
Safety Approvals Input Voltage Range
100~240VAC
Vin
Application:
Operate Voltage Range
90~264VAC
47~63Hz
- Blood Pressure system
- Portable medical devices
- ECG machine
Input Frequency
fin
Output Power Range
Output Voltage Range
Output Current Range
Input Current (Low Line)
Input Current (High Line)
See rating chart
See rating chart
See rating chart
Po
Vo
Io
Io=Full load, Vin=100VAC
Io=Full load, Vin=240VAC
Io=Full load, 25°C, Cool start, Vin=115VAC
Io=Full load, 25°C, Cool start, Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
0.6A
Iil
0.4A
Iih
28A (max.)
56A (max.)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Ir
Safety Approvals:
73.5~85%
1% (max.)
5% (max.)
Eff
Line Regulation
REG-i
REG-o
OVP
EN 60601-1
IEC 60601-1
P
S
RT
Load Regulation
Vin=230VAC
E
Over Voltage Protection
112~132%
110~150%
4mS (max.)
12mS (min.)
RoHS
2002/95/EC
COMPLIANT
OCP Over Current Protection
Time of Transient Response
Hold-Up Time
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Ttr
Th
Ts
Start Up Time
0.3~2S
Vp-p Ripple & Noise(Peak to Peak)
1% (max.)
All output
Temperature Coefficient
TC
0.04%/°C
Operating Altitude
Up to 3000m
See rating chart
6284VDC (min.)
OA
Pno
Vps
No-Load Power Consumption
Dielectric Withstanding Voltage
No load, Vin=230VAC
Primary to secondary
Note: The specifics for testing the energy efficiency of TPHPU32 Series are outlined in a separate document titled
"Test Method for Calculating the Energy Efficiency of Single-Voltage External Ac-Dc and Ac-Ac Power Supplies
(August 11, 2004)," which is available on the ENERGY STAR Website.
Environmental
Operating Temperature
See derating crve
-40~85°C
To
Ts
Storage Temperature
Operating Humidity
0~95%
Ho
Storage Humidity
0~95%
Hr
Operating Temperature at 25°C, Calculated per MIL-HDBK-217F
Derate linearly from 100% load at 40°C to 50% load at 70°C
0.1M Hrs (min.)
MTBF
Pd
Total Power International, Inc. Toll Free: 877-646-0900(U.S.) PH# (978) 453-7272
www.total-power.com sales@total-power.com