Total Power International, Inc.
15W Interchangeable Power Supplies for MEDICAL
TPHPU15 SERIES
Features:
Wide Operating Voltage, 80 to 275 VAC, 47 to 440 Hz
Interchangeable Plug
Single Output
Over Voltage and Over Load protection
Medical Safety 3rd (IEC60601-1 3rd Edition)
Input to Output : 2MOPP
Energy Star 2.0, Efficiency level V
Class II system
2 year warranty
Electrical Characteristics:
Safety Approvals Input Voltage Range Safety Approval & Specification in Label
Vins
Vin
Fi
100~240VAC
80~275VAC
47~440Hz
15W
Application:
- Home Healthcare Equipment
- Patient Monitor
Operate Voltage Range
Input Frequency
See derating curve
Sine wave
Output Power Range
Input Current (Low Line)
Input Current (High Line)
See rating chart
Po
Iil
- Blood Pressure system
- Portable medical devices
- ECG machine
Io=Full load, Vin=100VAC
Io=Full load, Vin=240VAC
Io=Full load, 25°C, Cool start, Vin=100VAC
Io=Full load, 25°C, Cool start, Vin=240VAC
Io=Full Load, Vin=230VAC
Io=Full Load, Vin=100~240VAC
Vin=230VAC
0.6A
0.4A
Iih
35A (max.)
85A (max.)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Ir
68~85%
1% (max.)
3~5%
Eff
Safety Approvals:
CB
IP22
Line Regulation
REG-i
REG-o
OVP
Load Regulation
Over Voltage Protection
Nil
Nil.But,Output protected to short circuit conditions
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100~240VAC
Full Load, Vin=90VAC
OCP Over Current Protection
4mS (max.)
12mS (typ.)
2S (max.)
Time of Transient Response
Hold-Up Time
Ttr
Thu
Ts
RoHS2
S&E
2011/65/EU
Start Up Time
Vp-p Ripple & Noise(Peak to Peak)
1% (max.)
Temperature Coefficient
Tc
All Condition
0.04%/°C
See rating chart
6483VDC (min.)
No-Load Power Consumption
Dielectric Withstanding Voltage
No load, Vin=230VAC
Pno
Vps
Primary to secondary
Compliance to EN55011 (CISPR11),
EN61000-3-2,-3
B Class
EMI EMC Emission
Environmental
Operating Temperature
To
Ts
See derating curve
-10~70°C
-40~85°C
0~95% RH
0~95% RH
2kV (min.)
8kV (min.)
6kV (min.)
Storage Temperature
Operating Humidity
Storage Humidity
Surge Voltage
10 ~ 95% RH
Non-condensing
Ho
Hr
See Rating Chart
All Condition
Vsg
ESDa Electro Static Discharge
Air Discharge, IEC61000-4-2
Contact Discharge, IEC61000-4-2
Electro Static Discharge
ESDc
Operating Temperature at 25°C,
Calculated per MIL-HDBK-217F
0.1M Hrs (min.)
5000m (max.)
5G (min.)
MTBF Mean Time Between Failure
Operating Altitude (Elevation)
ELEV
All condition
10 ~ 500Hz, 10min./1cycle, 60min.
Each along X, Y, Z axes
VBR Vibration
Total Power International, Inc. Toll Free: 877-646-0900(U.S.) PH# (978) 453-7272
www.total-power.com sales@total-power.com