UNIVERSAL INPUT AC TO DC MEDICAL APPLICATION
EXTERNAL ADAPTER SINGLE OUTPUT 45-50 WATTS
SWITCHING POWER SUPPLIES TPES49 SERIES
FEATURES:
ꢀ ACCOMMODATE UNIVERSALAC SOURCES
ꢀ DESKTOP IEC320-2P/3PAC RECEPTACLE
ꢀ MEET MEDICAL STANDARDS EN60601/ UL2601
ꢀ EMI MEET EN 55022/ FCC CLASS B
ꢀ CE MARKING COMPLIANCE
SPECIFICATION
INPUT SPECIFICATION
Input Voltage: 90-264Vac typical.
Input Connector: IEC320-3P DT7 or 2P DT8
AC input receptacle.
OUTPUT SPECIFICATION
Output Voltage: See Ratings Chart.
Output Current: See Ratings Chart.
Output Wattage: 45-50Watts typical.
Output Connector & Cord: Optional.
Line Regulation: 0.1% typical.
Load Regulation: ±1.5-3.0% typical.
Noise & Ripple: Typ. 1.0% peak to peak.
OVP: Built-in by latch circuit.
Input Frequency: 47-63Hz.
Inrush Current: 32A @230Vac.
Input Current: 0.79A@115Vac & 0.37A @230Vac.
Dielectric Withstand: Meet IEC601.
EMI: Meet EN55022 / FCC Class B.
Hold-up Time: 15mS @115Vac.
60mS @230Vac.
Adjustability: Factory set.
Over Current Protection (OCP): Installed.
Current limiting can be set precisely as per request.
Over Temp. Protection: Optional (NTC ckt.).
Earth Leakage: Less than 0.1 mA.
GENERAL SPECIFICATION
Efficiency: 85-87% typical.
Operating Temperature: 0 to +40°C
Storage Temperature: -20 to +85°C.
Temperature Coefficient: 0.04% /°C
Cooling: Free air convection.
Construction: Impact resistant thermo-plastic
enclosure case.
Switching Frequency: Typ. 85 ~ 110K Hz.
Circuit Topology: Fixed Frequency flyback circuit.
Transient Response: Output voltage returns in
less than 3mS following a 25% load change.
Safety Standard: Medical standards IEC601/EN60601/
UL2601, Class I for DT7, or Class II for DT8 type.
Power Density: 3.3Watts. / Cubic Inch.
Desktop Format.
Note: (1) All measurements are at nominal input, full load, and +25°C unless otherwise specified.
(2) Load regulation is measured at 115Vac or 230Vac in percentage to indicate the change in output voltage as the load varied from half load to full load (±%).
(3) The exact obtainable load regulation depends upon the output cord selected and load current. Upper data are for 6 ft. (2 m)cord AWG#18 wires.
For the details of safety approval, please consult the factory.
TOTAL POWER INT'L Tel: 877-646-0900 Fax: 978-453-7395
ꢀ
Email: sales@total-power.com http://www.total-power.com
-1-
ꢀ