TMP92 Series
90 Watts Medical Grade PowerAdapters
FEATURES
• 0.87" Slim Case
• Low Cost
• Class I (2MOPP) & Class II (1MOPP) Models Available
• Low Safety Ground Leakage Current
• Compliant with CEC & Energy Star Level V
• 100% Burn-in Tested
INPUT SPECIFICATIONS
MODELS LIST
Input Voltage Range ............ 90~264 VAC
Input Frequency ................... 47~63Hz
Output
Voltage
Maximum
Current
Maximum
Power
Product No.
Input Current ........................ 1.5A rms @ 90 VAC
Inrush Current ..................... 40A @ 115 VAC or 80A @ 230 VAC,
Class I (IEC 320/C6 Inlet)
°
at 25 C cold start
TMP92-12
TMP92-18
TMP92-19
TMP92-24
12 V
18 V
19 V
24 V
7.5 A
5 A
90 W
90 W
90 W
90 W
Earth Leakage Current ........ 160µA max. @ 264 VAC, 63Hz
............................................. (for Class I models only)
Touch Current ...................... 40µAmax. @ 264 VAC, 63Hz
............................................. (for Class I and Class II models)
No Load Power.................... 0.21W typical
4.74 A
3.75 A
Class II (IEC 320/C8 Inlet)
OUTPUT SPECIFICATIONS
TMP92-12-2R
TMP92-18-2R
TMP92-19-2R
TMP92-24-2R
12 V
18 V
19 V
24 V
7.5 A
5 A
90 W
90 W
90 W
90 W
Output Power Ratings .......... See table
Tolerance ............................. ±5%
Ripple and Noise* ............... 300mV peak to peak max. at full load
Overvoltage Protection ........ 36 Vdc max., Latch off
Overcurrent Protection ........ 6Amax., Auto-recovery
Over-Temp. Protection ........ Latch off
4.74 A
3.75 A
Short Circuit Protection ....... Auto-recovery
°
Temperature Coefficient ..... ±0.04%/ C max.
STANDARDS & COMPLIANCES
Turn-On Delay ..................... 3 sec. @ 100 VAC, 60Hz
Transient Response........... Max. excursion of 4% or better on
all models, recovering to 1% of
IEC/EN 60601-1-2 ............. EMC & Immunity Performance
EN 55011, FCC, VCCI ....... Class B, conducted & radiated
EN 61000-3-2 .................... Harmonic distortion, Class A
EN 61000-3-3 .................... Line flicker
final value within 500 µs after a
25% step load change
EN 61000-4-2 .................... ESD, ±8 KV air and ±6 KV contact
EN 61000-4-3 .................... Radiated immunity, 3 V/m
EN 61000-4-4 .................... Fast transient/burst, ±2 KV
EN 61000-4-5 .................... Surge, ±1 KV diff., ±2 KV com.
EN 61000-4-6 .................... Conducted immunity, 3 Vrms
EN 61000-4-8 .................... Magnetic field immunity, 3 A/m
EN 61000-4-11 .................. Voltage dip immunity,
30% reduction for 500ms,
* Measured with 20MHz bandwidth at rated line voltage and output load ranges,
with a 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor in
parallel across the output.
GENERAL SPECIFICATIONS
Power Factor ....................... 0.9 typical @ 115 VAC
Efficiency ............................. 88% minimum at full load
Hold-up Time....................... 10 ms min. @ 115 VAC, 60Hz
Switching Frequency ........... 75~100KHz
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ............... IEC/EN/UL 60601-1 3rd Edition,
ANSI/AAMI ES 60601-1 1st Edition,
Line Regulation................... ±1% maximum at full load
°
°
Operating Temperature ...... 0 C to +40 C
°
Derating............................... Derate from 100% at +40 C linearly
°
CSA-C22.2 No. 60601-1 2nd Edition
AgencyApprovals .............. UL, TUV, CE, CB
Other Compliances .......... RoHS, CEC & Energy Star Level V
to 50% at +60 C
°
°
Storage Temperature .......... -40 C to +85 C
Relative Humidity ................ 5% to 90% non-condensing
Withstand Voltage ............... 4000 VAC from input to output
............................................. 1500 VAC from input to ground
MTBF .................................... 100K hours minimum at full load,
°
25 C ambient, calculated per
MIL-HDBK-217F
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com