SMDA15B SERIES
15W Desktop Power Supply For Medical Equipment
–
–
–
–
–
–
–
Wide Input Voltage 90 to 264 VAC, 47 to 63Hz
IEC-320-C8 input inlet
Single Output
Output Voltage Available From 5VDC Thru 36VDC
Over Voltage, Over Load, and Over Temperature Protection
Class II Insulation
Energy Star 2.0, CEC Level V, and RoHS Compliance
3 Year Warranty
Approvals:
Single Output
Part Number
*SMDA15B-S02
*SMDA15B-S03
*SMDA15B-S04
*SMDA15B-S05
SMDA15B-S06
Output Voltage
Max. Output Current
2.60 ~ 2.16 A
2.16 ~ 1.62 A
1.87 ~ 1.36 A
1.36 ~ 1.15 A
1.15 ~ 0.93 A
0.93 ~ 0.71 A
0.71 ~ 0.55 A
0.55 ~ 0.45 A
0.45 ~ 0.41 A
Total Regulation
Max. Output Power
5 ~ 6 VDC
6 ~ 8 VDC
5%
5%
5%
5%
5%
5%
3%
3%
3%
13W
13W
15W
15W
15W
15W
15W
15W
15W
8 ~ 11 VDC
11 ~ 13 VDC
13 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
33 ~ 36 VDC
SMDA15B-S07
SMDA15B-S08
SMDA15B-S09
SMDA15B-S10
Mark " * " means " PSE approval "
The total regulation on model S02, S03, S05 is required to use AWG#16 / 4FT output cable.
The total regulation on model S04, S06~S10 is required to use AWG#18 / 4FT output cable.
The regulation and efficiency are not guaranteed if changes the output cable
Electrical Characteristics
Parameter
Test Conditions
Min.
90
47
Typ.
Max.
264
63
Unit
VAC
Hz
W
A
A
A
A
%
Input Voltage
Operating Voltage
Input Frequency
Output Power Range
Input Current (Low Line)
Input Current (High Line)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Vin=90 to 264VAC
0
15
Io=Full load, Vin=115VAC
Io=Full load, Vin=230VAC
Io=Full load, 25°C Cool start, Vin=115VAC
Io=Full load, 25°C Cool start, Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
0.25
0.17
14
0.33
0.18
16
30
85
28
73
Line Regulation
Load Regulation
0.5
3
1
5
%
%
Vin=230VAC
1
Over Voltage Protection*
Over Current Protection*
Transient Response
Hold-Up Time
Start Up Time
Ripple & Noise(Peak to Peak)
Safety Ground Leakage Current
Temperature Coefficient
No-Load Power Consumption
Thermal Shutdown By Junction
Temperature Controller *
112
110
132
150
4
%
%
mS
mS
S
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Io=Full Load, Vin=240VAC
All output
No load, Vin=240VAC
The parameter is not subject to production
test-verified by design/characterization of integrated
controller. Auto recovery.
10
0.3
16
1
0.5
2
1
0.3
0.04
0.3
%
mA
%/°C
W
-0.04
-20
0.25
125
°C
*When the power system interruption is isolated, the product would re-start after recovering by hand.