SMDA15A SERIES
15W Desktop Power Supply For Medical Equipment
–
–
–
–
–
–
–
Wide Input Voltage 90 to 264 VAC, 47 to 63Hz
IEC-320-C14 input inlet
Single Output
Output Voltage Available From 5VDC Thru 36VDC
Over Voltage, Over Load, and Over Temperature Protection
Class I Insulation
Energy Star 2.0, CEC Level V, and RoHS Compliance
3 Year Warranty
Approvals:
Single Output
Part Number
SMDA15A-S02
SMDA15A-S03
SMDA15A-S04
SMDA15A-S05
SMDA15A-S06
SMDA15A-S07
SMDA15A-S08
SMDA15A-S09
SMDA15A-S10
Output Voltage
5 ~ 6 VDC
Max. Output Current
2.60 ~ 2.16 A
2.16 ~ 1.62 A
1.87 ~ 1.36 A
1.36 ~ 1.15 A
1.15 ~ 0.93 A
0.93 ~ 0.71 A
0.71 ~ 0.55 A
0.55 ~ 0.45 A
0.45 ~ 0.41 A
Total Regulation
Max. Output Power
5%
5%
5%
5%
5%
5%
3%
3%
3%
13W
13W
15W
15W
15W
15W
15W
15W
15W
6 ~ 8 VDC
8 ~ 11 VDC
11 ~ 13 VDC
13 ~ 16 VDC
16 ~ 21 VDC
21 ~ 27 VDC
27 ~ 33 VDC
33 ~ 36 VDC
The total regulation on model S02, S03, S05 is required to use AWG#16 / 4FT output cable.
The total regulation on model S04, S06~S10 is required to use AWG#18 / 4FT output cable.
The regulation and efficiency are not guaranteed if changes the output cable
Electrical Characteristics
Parameter
Test Conditions
Min.
90
47
Typ.
Max.
264
63
Unit
VAC
Hz
W
A
A
A
A
%
Input Voltage
Operating Voltage
Input Frequency
Output Power Range
Input Current (Low Line)
Input Current (High Line)
Low Line Inrush Current
High Line Inrush Current
Efficiency
Vin=90 to 264VAC
0
15
Io=Full load, Vin=115VAC
Io=Full load, Vin=230VAC
Io=Full load, 25°C Cool start, Vin=115VAC
Io=Full load, 25°C Cool start, Vin=230VAC
Io=Full Load, Vin=230VAC
Io=Full Load
0.25
0.17
14
0.33
0.18
16
30
85
28
73
Line Regulation
Load Regulation
0.5
3
1
5
%
%
Vin=230VAC
1
Over Voltage Protection*
Over Current Protection*
Transient Response
Hold-Up Time
Start Up Time
Ripple & Noise(Peak to Peak)
Safety Ground Leakage Current
Temperature Coefficient
No-Load Power Consumption
Thermal Shutdown By Junction
Temperature Controller *
112
110
132
150
4
%
%
mS
mS
S
Io=Full Load to Half Load, Vin=100VAC
Io=Full Load, Vin=110VAC
Io=Full Load, Vin=100VAC
Full Load, Vin=90VAC
Io=Full Load, Vin=240VAC
All output
No load, Vin=240VAC
The parameter is not subject to production
test-verified by design/characterization of integrated
controller. Auto recovery.
10
0.3
16
1
0.5
2
1
0.3
0.04
0.3
%
mA
%/°C
W
-0.04
-20
0.25
125
°C
*When the power system interruption is isolated, the product would re-start after recovering by hand.