PSIM360B SERIES
360 Watts
AC/DC Open Frame Medical Power Supply
Single Outputs
Rev C
FEATURES
• Both ITE and Medical Approvals
• High Power Density, 360W in 6” x 4”
Footprint
• Medical Applications Protection: Means of
Patient Protection (MOPP)
• 5Vsb Meets ErP 0.5w @ No Load
• Meets Medical BF Rating
• Main Output and Standby Output Power ON
LED Indicators
• Remote Control (Inhibit) Function
• RoHS Compliant
• Wide Input Range of 90-264VAC
• Altitude During Operation: ITE up to 5000
m, Medical below 3000M
DESCRIPTION
The PSIM360B series of AC DC open frame medical power supplies offers 360 watts of
output power in a highly reliable 6" x 4" x 1.18" frame. This series consists of single
output models, input range of 90-264VAC, and a frequency of 47-63Hz. All models
have high efficiency above 90%, remote control (inhibit) function, and high power
density. This series meets medical BF rating as well as ITE and medical approvals.
Please call factory for order details.
Size: 6in x 4in x 1.18in
MODEL SELECTION TABLE
Maximum Load (V1)
Output Power(3)
Standby
Supply
(V2)
Fan
Input Voltage Output
Ripple &
Output
20CFM
Forced Air
Total Power
Model Number(1)
Output
Efficiency
>90%
18 CFM
Forced Air
Convection
Total Power
Range
Voltage
Noise(4) Regulation
Convection
(V3)(2)
12V
24V
28V
48V
54V
20.84A
10.42A
8.93A
5.21A
4.63A
30A
15A
120mV
240mV
280mV
300mV
300mV
±3%
±3%
±2%
±2%
±2%
250W
250W
250W
250W
250W
360W
360W
360W
360W
360W
5V/0.5A 12V/0.3A
5V/0.5A 12V/0.3A
5V/0.5A 12V/0.3A
5V/0.5A 12V/0.3A
5V/0.5A 12V/0.3A
PSIM360B-1Y120Z
PSIM360B-1Y240Z
PSIM360B-1Y280Z
PSIM360B-1Y480Z
PSIM360B-1Y540Z
90-264VAC
12.86A
7.5A
6.67A
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
INPUT SPECIFICATIONS
Input Voltage Range
Input Frequency
See derating curve
90
47
264
63
VAC
Hz
@115VAC
@230VAC @Full Load
@115VAC
@230VAC cold start @25ºC
@115VAC
@230VAC max
@264VAC
>0.95
>0.90
<40
<80
5
Power Factor
Inrush Current
Input Current (rms)
A Peak
A
2.5
<100
Earth Leakage Current
OUTPUT SPECIFICATIONS
Output Voltage
uA
See Table
Output Current
Remote Control
See Table
Inhibit function
360
20CFM Forced Air
Convection Cooling
@Full Load, 115VAC
Output Power
W
250
Hold Up Time
>10
ms
Ripple & Noise (20MHz bandwidth)
PROTECTION
See Table
Short Circuit Protection
Auto-Recovery
150
% Max.
Rating
Over Power Protection
Auto-Recovery
105
Over Voltage Protection
Latching Type. AC Recycle
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Humidity
See derating curve
0
-20
0
70
85
90
ºC
ºC
%
MTBF
@Full load and 25ºC ambient temperature per Telcordia (Bellcore TR-332)
>250,000
hours
Wall Industries, Inc. • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
website: www.wallindustries.com • e-mail: sales@wallindustries.com
9/15/2021