PM100 Series
3 7 - 6 4 W AT T S - A C / D C S I N G L E O U T P U T
FEATURES
•
•
•
•
•
•
•
Medical and ITE approvals
Compact size
High power density 10 W/cubic inch
100 W output with convection cooling up to +50℃
Low earth leakage current
EN55011 /55022 class B emissions
RoHS compliant
SPECIFICATIONS
INPUT
OPERATING
Voltage range
90-132 /180-264 VAC (Universal mains
supply operation)
MTBF
270,000 hours at full load at 25℃
ambient temperature, calculated per MIL-
HDBK-217F
Frequency
47-63 Hz
Efficiency
88-90% @ 230 VAC full load
100 KHz (typical)
Inrush current
40 A @ 115 VAC or 80 A @ 230 VAC, at
25°C cold start
1.9 A (rms) for 100-120 VAC
1.1 A (rms) for 200-240 VAC
Switching frequency
ENVIRONMENT
Input current
Operating temperature:
Storage temperature:
-10°C to +70°C
Earth leakage current
Isolation
150 µA max. @ 264 VAC, 63 Hz
-40°C to +85°C
Input – Output: 4000VAC
Input – Case:1500VAC
Output – Case: 500VAC
Relative humidity:
Derating:
5% to 95% non-condensing
Derate from 100% to +50°C linearly to
50% at +70°C
OUTPUT
Output voltage
See table
See table
STANDARDSANDAPPROVALS
Safety
UL 60601-1, CSA C22.2 No. 601.1, TÜV EN
Output current
60601-1,
60950-1, TÜV EN 60950-1
UL 60950-1, CSA C22.2 No.
Ripple and noise
150 mV peak to peak on 5.0 V model,
1% peak to peak on other models
C-Tick
AS/NZS CISPR11 Group1, Class A
Line regulation
0.2% maximum at full load
See table
EMC standards
IEC60601-1-2, EN55011 / EN55022 and
FCC, VCCI: Class B conducted . class B
radiated
Load regulation
Transient response
Maximum excursion of 4% or better on
all models, recovering to 1% of final
value within 500 us after a 25% step
load change
Immunity standards
EN61000-3-2, EN61000-3-3, EN61000-4-
2, EN61000-4-3, EN61000-4-4, EN61000-
4-5, EN61000-4-6, EN61000-4-8,
EN61000-4-11
Overvoltage protection
Overcurrent protection
Provided on output; set at 110-140% of
itsnominal output voltage
All outputs protected to short circuit
conditions
MECHANICAL
Dimensions
Weight
101x50.8x32mm
190g
Temperature coefficient
Hold-up time
All outputs 0.04% /°C maximum
12 ms minimum at 110 VAC
Doc. No. PBC9 Rev.C 28-4-11
400