MPU30 SERIES
30W Desk Top Switching Power Supplies For Medical Equipment.
Description:
The MPU30 series of AC/DC switching mode power supplies provide
30 Watts of continuous output power . All supplies are UL94V-1 min
compliant, include IEC-320-C14 input for worldwide applications.
They are ideally suited for use in hospital instrument and many
other applications. All models meet FCC Part-18 class B and CISPR-
11 EN55011 class B emission Limits and are designed to comply
with UL/c-UL(UL 60601-1) ,TUV/T-mark (EN 60601-1) and new CE
requirements. All units are 100% burned in and tested.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single to Triple Output
Output Voltage Available From 3VDC Thru 40VDC
Optional Output Connector (See appendix)
Input Surge Current, Over Voltage And Over Load protection
Splash proof
Output Voltage Protection(Crowbar Design)
Safety Approvals :
Class I Insulation
2 Year Warranty
Electrical Characteristics:
Sym.
Vin
Fin
Wo
Test Conditions
Typ.
Max. Unit
Parameter
Min.
Input Voltage
Operating Voltage
VAC
Hz
W
264
63
90
Input Frequency
47
Output Power Range
30
Vin=90 to 264VAC
0
Output Voltage Range
Output Current Range
Input Current(Low Line)
Input Current(High Line)
See rating chart
Vo
V
Io
See rating chart
A
Iil
Iih
Io=Full load, Vin=115 VAC
Io=Full load,Vin=230 VAC
0.9
A
0.34
A
load, 25 C,Coolstart,
Io=Full
A
Low Line Inrush Current
Irl
12
15
Vin=115VAC
Io=Full
load, 25 C,Coolstart,
High Line Inrush Current
Efficiency
Regulation
26
30
A
Irh
Vin=230VAC
Io=Full Load, Vin=230VAC
Eff
REG-i
REG-o
OVP
75
83
0.5
2
88
1
%
%
%
Line
Io=Full Load
Vin=230VAC
Load Regulation
10
Over Voltage Protection
112
110
132
150
4
%
%
OCP
Over Current Protection
mS
mS
Ttr Transient Response
Io=Full Load to
Load,Vin=100VAC
Half
Io=Full Load, Vin=110VAC
Load,
Hold-Up Time
Thold
12
Ts
S
%
0.3 1.5
1
2
2
Start Up Time
Io=Full
Vin=100VAC
Vrn
Ilk
Tc
Ripple & Noise (Peak toPeak)
Safety Ground Leakage Current
Temperature Coefficient
Full Load, Vin=90VAC
Io=FullLoad, Vin=240VAC
All output
0.1
0.04
mA
-0.04
%/ C
Environmental :
Sym.
Toper
Tstg
Hr
Parameter
Conditions
Test
Unit
Min. Typ.
Max.
70
Operating Temperature
Storage Temperature
0
-40
5
C
C
%
85
Humidity
Relative
95
Derate linearly from 100% load at 50 C to 50% load at 70 C
Pd
Safety Specifications:
Parameter
Test Conditions
Min. Typ. Max. Unit
Sym.
Dielectric Withstanding Voltage
for Primary to secondary
Vps
VDC
Primary to secondary
5600
Dielectric Withstanding Voltage
for Primary to Ground
Vpg
Ri
VDC
2800
Primary to ground
Isolation Resistance
Test Voltage=2100VDC
Vin=220VAC
M
50
CISPR EMIrequirements for CISPR-11
for FCC PART-18
FCC EMIrequirements
B
CLASS
CLASS
B
Vin=110VAC