MPU100 SERIES
100W Desk Top Switching Power Supplies For Medical Equipment.
Description:
The MPU100 series of AC/DC switching mode power supplies
provide 100 Watts of continuous output power . All supplies are
UL94V-1 min compliant, include IEC-320-C14 input for worldwide
applications. They are ideally suited for use in hospital instrument
and many other applications. All models meet FCC Part-18 class B
and CISPR-11 EN55011 class B emission Limits and are designed
to comply with UL/c-UL(UL 60601-1) ,TUV/T-mark (EN 60601-1)
and new CE requirements. All units are 100% burned in and tested.
Features:
Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
IEC-320-C14 Input Inlet
Single to Triple Output
Output Voltage Available From 5VDC Thru 40VDC
Optional Output Connector (See appendix)
Input Surge Current, Over Voltage And Over Load protection
Splash proof
Power Factor Correction
Output Voltage Protection(Crowbar Design)
Safety Approvals :
Class I Insulation
ON/OFF SWITCH
Two Years Warranty
Electrical Characteristics:
Sym.
Parameter
Test Conditions
Min. Typ. Max. Unit
Vin
Fin
Wo
Input Voltage
Operating Voltage
264
63
90
47
0
VAC
Hz
W
Input Frequency
Output Power Range
100
Vin=90 to 264VAC
Output Voltage Range
Output Current Range
Input Current(Low Line)
Input Current(High Line)
See rating chart
See rating chart
1.25
Vo
V
Io
A
Iil
Iih
Io=Full load, Vin=115 VAC
Io=Full load,Vin=230 VAC
A
0.5
A
load, 25 C,Coolstart,
Io=Full
A
Low Line Inrush Current
Irl
12
15
Vin=115VAC
Io=Full
load, 25 C,Coolstart,
High Line Inrush Current
Efficiency
Regulation
26
30
A
Irh
Vin=230VAC
Io=Full Load, Vin=230VAC
Eff
75
83
0.5
3
88
1
%
%
%
REG-i
REG-o
Line
Io=Full Load
Vin=230VAC
Load Regulation
7
Over Voltage Protection
112
110
132
150
4
OVP
OCP
%
%
Over Current Protection
Ttr Transient Response
mS
mS
Io=Full Load to
Load,Vin=100VAC
Half
Hold-Up Time
Thold
Io=Full
Load,
Vin=110VAC
16
Load,
Start Up Time
Ts
Io=Full
Vin=100VAC
0.3 1.5
1
2
2
S
%
Vrn
Ilk
Tc
Ripple & Noise (Peak toPeak)
Safety Ground Leakage Current
Temperature Coefficient
Full Load, Vin=90VAC
Io=FullLoad, Vin=240VAC/60Hz
All output
mA
0.1
-0.04
0.04
%/ C
Environmental :
Sym.
Toper
Tstg
Hr
Parameter
Conditions
Test
Unit
Min. Typ.
Max.
70
Operating Temperature
Storage Temperature
0
-40
5
C
C
%
85
Humidity
Relative
95
Derate linearly from 100% load at 50 C to 50% load at 70 C
Pd
Safety Specifications:
Parameter
Test Conditions
Min. Typ. Max. Unit
Sym.
Dielectric Withstanding Voltage
for Primary to secondary
Vps
VDC
Primary to secondary
5600
Dielectric Withstanding Voltage
for Primary to Ground
Vpg
Ri
VDC
2800
Primary to ground
Isolation Resistance
M
50
Test Voltage=2100VDC
Vin=220VAC
CISPR EMIrequirements for CISPR-11
for FCC PART-18
FCC EMIrequirements
B
CLASS
CLASS
B
Vin=110VAC