1~4 Output Medical Type
120W Medical series
Features
Universal AC input / Full range
Low leakage current <180 A
Protections: Short circuit / Overload / Over voltage
UL60601-1 medical safety approved
With power good and fail signal output (Optional)
100% full load burn-in test
Fixed switching frequency at 45KHz
3 years warranty
EN60601-1
IEC60601-1
Specification
AC INPUT VOLTAGE
90~264 VAC, 47~440Hz / 127~370VDC.
TEMP. COEFFICIENT
0.04%/ on all outputs at full load between 0~50 of ambient temperature.
AC INPUT CURRENT (Typ.)
Maximum input current 2.9A at 115VAC, 60Hz or
1.7A at 230VAC, 60Hz with 100% output load.
VIBRATION
2G of acceleration, vibrating frequency adjust from 10Hz ~500Hz within a
10-minute cycle. 6 testing cycles (60 minutes) each along X, Y, Z axes.
INRUSH CURRENT (Typ.)
SAFETY STANDARDS
Medical : UL60601-1, TUV EN60601-1, IEC60601-1 approved
Commercial : Also design refer to UL60950-1, TUV EN60950-1
Inrush current is less than 22A at 115VAC or less than
45A at 230VAC under cold start conditions. Limiting
provided by internal thermistors.
WITHSTAND VOLTAGE
SETUP, RISE TIME
MPS-120:
4000VAC between input and output
1500VAC between input and F.G.
500VAC between output and F.G.
800ms, 20ms / 230VAC at full load
2000ms, 50ms / 115VAC at full load
MPD,T,Q-120:500ms, 20ms / 230VAC at full load
1200ms, 50ms / 115VAC at full load
ISOLATION RESISTANCE
>100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage.
HOLD-UP TIME (Typ.)
80ms / 230VAC at full load
14ms / 115VAC at full load
EMI COMPLIANCE
EMI Specifications
Conducted & Radiation
Compliance Level
EN55011, Class B
EN55022, Class B
EN61000-3-2
LEAKAGE CURRENT
Leakage current is less than 180 A at 264VAC
Harmonic distortion
Voltage flicker
EN61000-3-3
DC OUTPUT ADJ. RANGE
DC output voltage (or Ch1 of multiple output models)
can be adjusted between -5%~+10% rated output
voltage by potential meter.
EMS COMPLIANCE
EMS Specification
ESD air
ESD contact
RF field susceptibility
Compliance Level
EN61000-4-2, Level 3, 8KV
EN61000-4-2, Level 2, 4KV
EN61000-4-3, Level 2, 3V/m
Level 3, 10V/m
EN61000-4-4, Level 2, 1KV/5KHz
Level 3, 2KV/5KHz
OVERLOAD PROTECTION
Fully protected against short circuit and output overload.
The hiccup type protection will be activated at 110~150%
(For MPD,T,Q-120),120~160%(For MPS-120) rated load
and recovers automatically after fault condition is removed.
EFT(Electrical Fast Transient)/Burst
Lightning/Surge
EN61000-4-5, Level 4, 2KV/Line-Line
4KV/Line-Earth
OVER VOLTAGE PROTECTION
EN61000-4-6, Level 2, 3Vrms/m
Level 3, 10Vrms/m
EN61000-4-8, Level 2, 3A/m
Level 3, 10A/m
Conducted RF susceptibility
Magnetic field immunity
Provided on output channel 1 only at 115%~135%
rated output voltage. (120%~140% for MPS-120-15/24/48).
Output will be shut down when this protection is activated.
EN61000-4-11,Compliance
ENV50204, Level 2, 3V/m, 900MHz
Level 3, 10A/m, 900MHz
Voltage dip, interruption
Digital phone carrier immunity
POWER GOOD / FAIL SIGNAL (OPTIONAL)
TTL logic high for power good and TTL low for
power fail. When the output voltage reaches 90% of
rated value, a +5V TTL signal will be sent out with a
10~500ms delay; At least 1ms before the output
voltage goes below 90% of the rated value, the TTL
signal will be turned off.
MTBF
262,100 hours min. at full load and 25 of ambient temperature,
calculated per MIL-HDBK-217F.
DIMENSION (L*W*H)
177.8x107.95x35.5mm or 7"x4.25"x1.4"
* MPS-120-3.3 does not have this optional function.
WORKING TEMP.
Whole series can operate from -20~70 . Please refer
to the derating curves.
PACKING
0.55Kg; 24pcs/14.5Kg/1.04CUFT
WORKING HUMIDITY
20~90% RH non-condensing.
STORAGE TEMP., HUMIDITY
-40~+85 , 10~90% RH
File Name:120W MEDICAL-SPEC 2008-03-25