1~4 Output Medical Type  
120W Medical series  
Features  
Universal AC input / Full range  
Low leakage current <180 A  
Protections:Short circuit/Over load/Over voltage  
UL60601-1 medical safety approved  
With power good and fail signal output (Optional)  
100% full load burn-in test  
Fixed switching frequency at 45KHz  
3 years warranty  
IEC60601-1  
Specification  
AC INPUT VOLTAGE  
90~264 VAC, 47~440Hz / 127~370VDC.  
TEMP. COEFFICIENT  
0.04%/ on all outputs at full load between 0~50 of ambient temperature.  
AC INPUT CURRENT (Typ.)  
Maximum input current 2.9A at 115VAC, 60Hz or  
1.7A at 230VAC, 60Hz with 100% output load.  
VIBRATION  
2G of acceleration, vibrating frequency adjust from 10Hz ~500Hz within a  
10-minute cycle. 6 testing cycles (60 minutes) each along X, Y, Z axes.  
INRUSH CURRENT (Typ.)  
SAFETY STANDARDS  
Medical : UL60601-1, TUV EN60601-1 approved  
Commercial : Also design refer to UL60950-1, TUV EN60950-1  
Inrush current is less than 22A at 115VAC or less than  
45A at 230VAC under cold start conditions. Limiting  
provided by internal thermistors.  
WITHSTAND VOLTAGE  
SETUP, RISE TIME  
MPS-120:  
4000VAC between input and output  
1500VAC between input and F.G.  
500VAC between output and F.G.  
800ms, 20ms / 230VAC at full load  
2000ms, 50ms / 115VAC at full load  
MPD,T,Q-120:500ms, 20ms / 230VAC at full load  
1200ms, 50ms / 115VAC at full load  
ISOLATION RESISTANCE  
>100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage.  
HOLD-UP TIME (Typ.)  
80ms / 230VAC at full load  
14ms / 115VAC at full load  
EMI COMPLIANCE  
EMI Specifications  
Conducted & Radiation  
Compliance Level  
EN55011, Class B  
EN55022, Class B  
EN61000-3-2  
LEAKAGE CURRENT  
Leakage current is less than 180 A at 264VAC  
Harmonic distortion  
Voltage flicker  
EN61000-3-3  
DC OUTPUT ADJ. RANGE  
DC output voltage (or Ch1 of multiple output models)  
can be adjusted between -5%~+10% rated output  
voltage by potential meter.  
EMS COMPLIANCE  
EMS Specification  
ESD air  
ESD contact  
RF field susceptibility  
Compliance Level  
EN61000-4-2, Level 3, 8KV  
EN61000-4-2, Level 2, 4KV  
EN61000-4-3, Level 2, 3V/m  
Level 3, 10V/m  
EN61000-4-4, Level 2, 1KV/5KHz  
Level 3, 2KV/5KHz  
OVER LOAD PROTECTION  
Fully protected against short circuit and output overload.  
The hiccup type protection will be activated at 110~150%  
(For MPD,T,Q-120),120~160%(For MPS-120) rated load  
and recovers automatically after fault condition is removed.  
EFT(Electrical Fast Transient)/Burst  
Lightning/Surge  
EN61000-4-5, Level 4, 2KV/Line-Line  
4KV/Line-Earth  
OVER VOLTAGE PROTECTION  
EN61000-4-6, Level 2, 3Vrms/m  
Level 3, 10Vrms/m  
EN61000-4-8, Level 2, 3A/m  
Level 3, 10A/m  
Conducted RF susceptibility  
Magnetic field immunity  
Provided on output channel 1 only at 115%~135%  
rated output voltage. (120%~140% for MPS-120-15/24/48).  
Output will be shut down when this protection is activated.  
EN61000-4-11,Compliance  
ENV50204, Level 2, 3V/m, 900MHz  
Level 3, 10A/m, 900MHz  
Voltage dip, interruption  
Digital phone carrier immunity  
POWER GOOD / FAIL SIGNAL (OPTIONAL)  
TTL logic high for power good and TTL low for  
power fail. When the output voltage reaches 90% of  
rated value, a +5V TTL signal will be sent out with a  
10~500ms delay; At least 1ms before the output  
voltage goes below 90% of the rated value, the TTL  
signal will be turned off.  
MTBF  
262,100 hours min. at full load and 25 of ambient temperature,  
calculated per MIL-HDBK-217F.  
DIMENSION (L*W*H)  
177.8x107.95x35.5mm or 7"x4.25"x1.4"  
* MPS-120-3.3 does not have this optional function.  
WORKING TEMP.  
Whole series can operate from -20~70 . Please refer  
to the derating curves.  
PACKING  
0.55Kg; 24pcs/14.5Kg/1.04CUFT  
WORKING HUMIDITY  
20~90% RH non-condensing.  
STORAGE TEMP., HUMIDITY  
-40~+85 , 10~90% RH  
File Name:120W MEDICAL-SPEC 2006-04-13  

与MPT-120A相关器件

型号 品牌 获取价格 描述 数据表
MPT-120B ASTRODYNE

获取价格

120W 1~4 Output for Medical Type
MPT-120B XPPOWER

获取价格

1~4 Output Medical Type
MPT-120B MEANWELL

获取价格

1~4 Output Medical Type
MPT-120C ASTRODYNE

获取价格

120W 1~4 Output for Medical Type
MPT-120C XPPOWER

获取价格

1~4 Output Medical Type
MPT-120C MEANWELL

获取价格

1~4 Output Medical Type
MPT-120D ASTRODYNE

获取价格

120W 1~4 Output for Medical Type
MPT-120D MEANWELL

获取价格

1~4 Output Medical Type
MPT-120D XPPOWER

获取价格

1~4 Output Medical Type
MPT-12C MICROSEMI

获取价格

TRANSIENT ABSORPTION ZENER