mps/d/t/q-120 ꢀerieꢀ
120W 1~4 Output for Medical Type
Features
Universal AC�input�/�Full�range
Low�leakage�current�<180 A
Protections:Short�circuit/Over�load/Over�voltage
UL2601-1�medical�safety approved
With�power�good�and�fail�signal�output�(Optional)
100%�full�load�burn-in�test
Fixed�switching�frequency�at�45KHz
3�years�warranty
Specification
AC�INPUT�VOLTAGE
90~264�VAC,�47~440Hz�/�127~370VDC.
TEMP.�COEFFICIENT
0.04%/ on�all�outputs�at�full�load�between�0~50 of�ambient�temperature.
AC�INPUT�CURRENT (Typ.)
Maximum�input�current�2.9A at�115VAC,�60Hz�or
1.7A at�230VAC,�60Hz�with�100%�output�load.
VIBRATION
2G�of�acceleration,�vibrating�frequency�adjust�from�10Hz�~500Hz�within�a
10-minute�cycle.�6�testing�cycles�(60�minutes)�each�along�X, Y,�Z�axes.
INRUSH�CURRENT (Typ.)
SAFETY STANDARDS
Medical : UL2601-1, TUV�EN60601-1�approved
Commercial�: Also�design�refer�to�UL60950-1, TUV�EN60950-1
Inrush�current�is�less�than�22A at�115VAC�or�less�than
45A at�230VAC�under�cold�start�conditions.�Limiting
provided�by�internal�thermistors.
WITHSTAND�VOLTAGE
SETUP,�RISE�TIME
MPS-120:
4000VAC�between�input�and�output
1500VAC�between�input�and�F.G.
500VAC�between�output�and�F.G.
800ms,�20ms�/�230VAC�at�full�load
2000ms,�50ms�/�115VAC�at�full�load
MPD,T,Q-120:500ms,�20ms�/�230VAC�at�full�load
1200ms,�50ms�/�115VAC�at�full�load
ISOLATION�RESISTANCE
>100M�Ohms�for�I/P-O/P,�I/P-FG,�O/P-FG�by�using�500VDC�test�voltage.
HOLD-UP TIME (Typ.)
80ms�/�230VAC�at�full�load
16ms�/�115VAC�at�full�load
EMI�COMPLIANCE
EMI�Specifications
Conducted�&�Radiation
Compliance�Level
EN55011,�Class�B
EN55022,�Class�B
EN61000-3-2
LEAKAGE�CURRENT
Leakage�current�is�less�than�180 A at�264VAC
Harmonic�distortion
Voltage�flicker
EN61000-3-3
DC�OUTPUT ADJ.�RANGE
DC�output�voltage�(or�Ch1�of�multiple�output�models)
can�be�adjusted�between�-5%~+10%�rated�output
voltage�by�potential�meter.
EMS�COMPLIANCE
EMS�Specification
ESD�air
ESD�contact
RF�field�susceptibility
Compliance�Level
EN61000-4-2, Level�3,�8KV
EN61000-4-2, Level�2,�4KV
EN61000-4-3, Level�2,�3V/m
Level�3,�10V/m
EN61000-4-4, Level�2,�1KV/5KHz
Level�3,�2KV/5KHz
OVER�LOAD�PROTECTION
Fully�protected�against�short�circuit�and�output�overload.
The�hiccup�type�protection will�be�activated�at�110~150%
(For�MPD,T,Q-120),120~160%(For�MPS-120)�rated�load
and�recovers�automatically�after�fault�condition�is removed.
EFT(Electrical�Fast Transient)/Burst
Lightning/Surge
EN61000-4-5, Level�4,�2KV/Line-Line
4KV/Line-Earth
OVER�VOLTAGE�PROTECTION
EN61000-4-6, Level�2,�3Vrms/m
Level�3,�10Vrms/m
EN61000-4-8, Level�2,�3A/m
Level�3,�10A/m
Conducted�RF�susceptibility
Magnetic�field�immunity
Provided�on�output�channel�1�only�at�115%~135%
rated�output�voltage.�(120%~140%�for�MPS-120-15/24/48).
Output�will�be�shut�down�when�this�protection�is�activated.
EN61000-4-11,Compliance
ENV50204, Level�2,�3V/m,�900MHz
Level�3,�10A/m,�900MHz
Voltage�dip,�interruption
Digital�phone�carrier�immunity
POWER�GOOD�/�FAIL SIGNAL (OPTIONAL)
TTL logic�high�for�power�good�and TTL low�for
power�fail.�When�the�output�voltage�reaches�90%�of
rated�value,�a�+5V TTL signal�will�be�sent�out�with�a
10~500ms�delay; At�least�1ms�before�the�output
voltage�goes�below�90%�of�the�rated�value,�the TTL
signal�will�be�turned�off.
MTBF
262,100�hours�min.�at�full�load�and�25 of�ambient��temperature,
calculated�per�MIL-HDBK-217F.
DIMENSION�(L*W*H)
177.8x107.95x35.5mm�or�7"x4.25"x1.4"
*�MPS-120-3.3�does�not�have�this�optional�function.