MicroPower Direct
Go Direct - MicroPower Direct - For Product, Performance and Price
400W AC-DC ACTIVE PFC MEDICAL POWER SUPPLY
MPA400MD DUAL OUTPUT SERIES
FEATURES:
•
Medical Safety Approvals
• Power Factor Corrected to EN61000-3-2 class D
•
1U height size
•
U-Chassis & Enclosed with built-in fan Mechanical Options
•
High power density: 6.25 watts/cu inches
•
Current Monitoring and Remote Voltage adjustment (Margin)
U - Chassis Type: 8 x 5 x 1.6”
Enclosed with built-in fan Type: 9 x 5 x 1.6”
CB
PRODUCT SPECIFICATIONS:
Input Voltage: 90-264 Vac full range, 47~63 Hz.
Input Current: 6.35A at 90 Vac full load.
Inrush Current: 35A Max @ 230 Vac with full load and cold start.
PFC: Active power factor correction meet EN61000-3-2 class D.
Fan Drive: 12 VDC/400 mA is available to drive an external fan.
Short Circuit Protection: Trip without damage and auto-recovery.
Switching Frequency: 30 kHz fixed frequency.
Operating Temperature: O°C to 70°C ambient, de-rating at 2.5%
per degree from +50°C to +70°C
Storage Temperature: -20 to 85°C.
Transient Response: Returns to within 1% in less than 2.5 mS for Operating Humidity: 5% to 90% RH, Non-condensing.
a 50% load change and the peak transient does not excess 5%.
Overshoot: Turn-on/off not exceed 5% over nominal voltage.
Efficiency: 75% minimum (Measuring at 230V and full load).
Turn On Delay: 1 second maximum at 120 Vac.
Hold Up Time: 20 mS min. at 80% of full load.
Adjustability: Output user adjustable ±5% minimum.
Remote Sense: Designated “RS+” and “RS-” on the CN3.
Storage Humidity: 5% to 95% RH, Non-condensing.
Vibration: Frequency 5 to 50 Hz, acceleration ±7.35 M/(SxS) on
X,Y and Z Axis.
Emissions: FCC Part 15, CISPR 22 class B, Conducted.
Safety Regulation: Approved to UL60950-1/ 60601-1, CSA
C22.2 No. 60950-1-03/ 601.1-M90, TUV EN60950-1/ 60601-1, CE
Mark (LVD) EN61204-3/ 60601-1-2/ 61000-3-2,3 & IEC61000-4
Remote On-Off: Designated as “RSW” on the CN3, requires a low Series Regulations and CB.
signal to inhibit output.
Leakage Current: 300 uA
Power Supply On: Green LED designated as “LED 1” on the PCB. HI-POT Test: 1500 Vac between input line and chassis (2 mA DC
LED display: Bi-color “LED1” emit green light on operating. Any
protection occurred or RSW applied low signal will emit orange.
Power Good: Designated as “PG” on the CN3 will go high 100-
cut off current); 4000 Vac between primary and secondary
windings; Primary to core 1500 Vac. All for 3 sec.
Grounding Test: Apply 40 A from ground pin to the earthed
500mS after regulation and goes low 1mS before loss of regulation. connection point. Maximum allowable resistance is 0.1ohm.
Input Circuit (primary): Two protected fuse inserted, shall be
blown only by an internal power fault.
Input Voltage Protection: Power shut down under 80 ±5 Vac,
and recovered over 86 Vac.
Warranty: 2 years.
MTBF: 100000 Hrs (according to MIL-HBK-217F) at 30? .
Cooling: Enclosed Type: Self cooled by internal ball bearing fan.
U-Chassis Type: Full rating under 22.95 CFM forced air flow.
Over-Power Protection: C.C. mode 110-140% and auto-recovery. Burn in: 45 +/- 5 degree C for 1 hour @230Vac with full load.
Over-Voltage Protection: Latching down will occur when output
Enclosure: U-Chassis type - 8(L) x 5(W) x 1.6(H) inches.
voltage exceed 130% and recycle AC input to reset.
Enclosed type - 9(L) x 5(W) x 1.6(H) inches.
Over Temp. Protection: Protected at ambient 85°C; Auto-recovery. Weight: U-Chassis Type – 1.3KG; Enclosed Type – 1.6KG;
MicroPower Direct
www.micropowerdirect.com
292 Page Street, Ste D Stoughton, MA 02072 • TEL: (781) 344-8226 • FAX: (781) 344-8481 • EMAIL: sales@micropowerdirect.com