ML100MRI4
Medical Approved
Compact SMT, 1W
DC/DC Converters
Electrical Specifications
Specifications typical @ +25°C, nominal input voltage & rated output current, unless otherwise noted. Specifications subject to change without notice.
Key Features:
Input
Parameter
• EN 60601 3RD Ed. Approved
• 1W Output Power
Conditions
Min.
Typ.
5.0
Max.
5.5
Units
5 VDC Input
4.5
• 4 kVrms Isolation
Input Voltage Range
12 VDC Input 10.8
24 VDC Input 21.6
Internal Capacitor
12.0
24.0
13.2
26.4
VDC
• Reinforced Insulation
• 2 x MOPP per EN 60601-1
3RD Edition & ANSI/AAMI
ES 60601-1
Input Filter
Output
Parameter
Output Voltage Accuracy
Output Voltage Balance
Line Regulation
Conditions
Min.
Typ.
1.0
0.1
1.2
Max.
3.0
1.0
1.5
10
Units
%
%
%
%
Dual Outputs, Balanced Loads
For VIN Change of 1%
IOUT = 10% to 100%
See Note 2
• 2 µA Max Leakage Current
• Compact SMT Case
• Single & Dual Outputs
Load Regulation
Ripple & Noise (20 MHz)
Temperature Coefficient
Output Short Circuit
General
100
0.02
mV P - P
%/°C
0.01
Continuos (Autorecovery)
Parameter
Conditions
Min.
60 Seconds 4,000
250 Vrms
Typ.
Max.
2.0
Units
Vrms
Isolation Voltage
Reinforced Insulation Working Voltage
Leakage Current
Isolation Resistance
Isolation Capacitance
Switching Frequency
EMI Characteristics, See Note 3
Parameter
240 VAC, 60 Hz
500 VDC
µA
G⍀
pF
10
100 kHz, 1V
20
80
50
100
kHz
Standard
Criteria
Level
EMC
EMS
Complies With EN 55011 4TH Edition
Complies With EN 60601-1-2
Environmental
Parameter
Conditions
Min.
-25
Typ.
Max.
+95
+105
+125
Units
°C
°C
Ambient
Case
Operating Temperature Range
Storage Temperature Range
Cooling
Humidity
-50
°C
Free Air Convection
RH, Non-condensing
95
%
Physical
Case Size
See Mechanical Drawing (Page 2)
Case Material
Weight
Non-Conductive Black Plastic (UL94-V0)
0.14 Oz (4.1g)
Reliability Specifications
Parameter
MTBF
MicroPower Direct
292 Page Street
Suite D
Stoughton, MA 02072
USA
Conditions
Min.
2.0
Typ.
Max.
Units
MHours
MIL HDBK 217F, 25°C, Gnd Benign
IEC/EN 60601-1, EN 60601-1 3RD Edition, 2xMOPP
Safety Standards
ANSI/AAMI ES 60601-1, 2xMOPP Recognition, (UL Certificate)
ANSI/AAMI ES 60601-1, CAN/CSA-C22.2 No.60601-1
IPC/JEDEC J-STD-020D.1 Level 2
T: (781) 344-8226
F: (781) 344-8481
E: sales@micropowerdirect.com
W: www.micropowerdirect.com
Moisture Sensitivity Level, See Note 4
Absolute Maximum Ratings
Parameter
Conditions
Min.
Typ.
Max.
9.0
18.0
30.0
245
Units
VDC
°C
5 VDC Input
Input Voltage Surge (1 Sec)
12 VDC Input
24 VDC Input
See Note 5
Peak Reflow Temperature
Caution: Exceeding Absolute Maximum Ratings may damage the module. These are not continuous operating ratings.
www.micropowerdirect.com