Features
Meets UL/EN/IEC60601-1-2, 4th edition for EMC*
Approved to EN/IEC/UL60601-1, 3rd edition
2 MOPP Input to Output Isolation
.
.
.
.
Meets DoE Efficiency Level VI Requirements
.
.
No load input power
Average Efficiency
.
.
Up to 40W of AC-DC Power
Universal Input 90-264Vac Input Range
.
Desktop and Wall-Plug versions
.
Meets EN55011/CISPR11, FCC Part 15.109 Class B
Conducted & Radiated Emissions, with 6db margin
.
.
.
.
IP22 Rated Enclosure
E-cap life of >8 years
>1,000,000 hours MTBF
3 Year Warranty
Description
A high performance AC to DC external power supply family designed for medical applications. The ME40A Medical Series
low power external AC-DC power supplies are approved to safety EN/IEC/UL60601-1, 3rd edition with isolation levels
which satisfy the 2 MOPP requirements and designed to UL/EN/IEC60601-1-2, 4th edition for EMC*. The ME40A Series
models will operate at universal input range of 90 to 264Vac over the wide temperature range of -20°C to +70°C,
delivering full rated output power up to +40°C and applicable output power derating at 70°C. These models are available
in desktop and wall-plug versions, include an IP22 rating per IEC60529 for the enclosure, and output cable terminated at
a variety of output connectors.
*Consult Factory for Table 9 compliance information.
Model Selection
Overvoltage
Trip Range
Output
Connector
Input
Configuration
Model
Output
Current
Output
Power
Ripple &
Noise¹
Line
Regulation
Load
Regulation
Number
Volts
ME40A0503F01
ME40A1203F01
ME40A1803F01
5.0V
5.00A
3.40A
2.22A
25W
40W
40W
100mV pk-pk
120mV pk-pk
180mV pk-pk
120%-150%
120%-150%
120%-150%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class I Desktop,
IEC60320 C14
Receptacle
12.0V
18.0V
ME40A2403F01
ME40A0503N01
ME40A1203N01
ME40A1803N01
ME40A2403N01
ME40A0503Q01
ME40A1203Q01
ME40A1803Q01
ME40A2403Q01
ME40A0503B01
ME40A1203B01
ME40A1803B01
ME40A2403B01
ME40A0503C01
ME40A1203C01
ME40A1803C01
ME40A2403C01
24.0V
5.0V
1.70A
5.00A
3.40A
2.22A
1.70A
5.00A
3.40A
2.22A
1.70A
5.00A
3.40A
2.22A
1.70A
5.00A
3.40A
2.22A
1.70A
40W
25W
40W
40W
40W
25W
40W
40W
40W
25W
40W
40W
40W
25W
40W
40W
40W
240mV pk-pk
100mV pk-pk
120mV pk-pk
180mV pk-pk
240mV pk-pk
100mV pk-pk
120mV pk-pk
180mV pk-pk
240mV pk-pk
100mV pk-pk
120mV pk-pk
180mV pk-pk
240mV pk-pk
100mV pk-pk
120mV pk-pk
180mV pk-pk
240mV pk-pk
120%-140%
120%-150%
120%-150%
120%-150%
120%-140%
120%-150%
120%-150%
120%-150%
120%-140%
120%-150%
120%-150%
120%-150%
120%-140%
120%-150%
120%-150%
120%-150%
120%-140%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C8
Receptacle
12.0V
18.0V
24.0V
5.0V
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±5%
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Desktop,
IEC60320 C18
Receptacle
12.0V
18.0V
24.0V
5.0V
Class II Wall-Plug,
Interchangeable
Blades²
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
12.0V
18.0V
24.0V
5.0V
2.5 x 5.5 x 9.5mm
Straight Barrel Type,
center positive
Class II Wall-Plug,
Fixed North
American Blades³
12.0V
18.0V
24.0V
Notes:
1. Measured at the output connector, with noise probe directly across output and load terminated with 0.1μF ceramic and 10μF low ESR capacitors.
2. Standard models are fitted with North American blades. Order blade kit KT-1027K for other blades (EU. UK, Australia)
3. For EU fixed blades, replace “C” in the model number with “M”, for UK blades, replace “C” with “G”, for Australia blades, replace “C” with “H”.
4. All specifications are typical at nominal input, full load, at 25°C ambient unless noted.
5. For Input Class I models: For AC GND connected to output common (-), insert a “B” in the part number where the “A” is located (TE40B1203F01).
www.slpower.com