Medical & ITE Dual Safety Standards
65W Medical / ITE Power Supply
HPA65A/C series
V1.3
GENERAL DESCRIPTION:
FEATURES:
* Input Voltage Range 70-264VAC * Extremely Low Leakage Current <0.1mA
The series provides a voltage range from 8V to 48V, delivering up to 65
watts of power, suitable for various applications and devices.
(70VAC input is suitable for
* Operating Altitude: 5000M
countries with low input voltage) * Built-In EMI Filter
* Operating Temperature: -40~70°C * High Surge Immunity
(with derating from 50°C)
* Suitable for BF Application
* Protection: OVP, OLP, OTP, SCP
* 2MOPP
* IP22
* Pass LPS
* Designed to Meet ISN & 60335
* 3-Year Warranty
HPA65A
C14 Type AC Inlet
OPTIONAL FEATURES:
* OVC III
* Withstand 297VAC Surge Input
for 2 Minutes
* IP54
APPLICATIONS:
* Medical Devices
* Consumer Electronics
* Office Electronics
* Industrial Equipment
* Telecommunication Devices
* Household Devices
HPA65C
Dimensions (L × W × H):
C6 Type AC Inlet
115 x 50 x 31mm
SAFETY STANDARDS:
GENERAL SPECIFICATION:
* Waterproof and Dustproof Degree: IP22, IP54
* Protection Classes: Class I
CB
* Safety: IEC 62368-1 Edition 2.0, IEC 62368-1 Edition 3.0,
EN 62368-1, UL62368-1, CAN/CSA-C22.2 NO.62368-1-14,
IEC 60601-1 Edition 3.2, EN60601-1,
5K
IP22 IP54
ANSI/AAMI ES60601-1: 2005 (R2012), CSA C22.2 NO. 60601-1:14
ELECTRICAL CHARACTERISTICS:
Typ. Max.
Unit
Characteristic
Condition
Min.
100
70
Nominal Input Voltage
Input Voltage Range
100 / 240VAC, Single Phase
240 VAC
264 VAC
Derate Linearly from 100% load at 90VAC to 70% load at 70VAC
Sine Wave
Input Frequency Range
Output Power
47
63
65
Hz
W
A
See Rating Chart
Low Line Input Current
High Line Input Current
Low Line Input Inrush Current
High Line Input Inrush Current
Earth Leakage Current
Average Efficiency
Full Load, Vin=115VAC
1.5
0.75
50
Full Load, Vin=230VAC
A
Full Load, 25°C, Cool Start, Vin=115VAC
Full Load, 25°C, Cool Start, Vin=230VAC
0.1mA @264VAC Normal Condition, 0.3mA @264VAC Single Fault Condition
CoC v5 (Tier2), See Rating Chart
Full Load, Vin=100~120VAC
A
100
0.1
A
mA
%
89
Line Regulation
0.5
5
%
Load Regulation
Vin=100~240VAC
%
3
Hold up time is measured from the end of the last charging pulse to the time which
the main output drops down to low limit of main output at rated load and nominal line.
Hold-Up Time
12
ms
Full Load, Vin=100~240VAC
At 115VAC & 230VAC
3
s
ms
Start-Up Time
Rise Time
50
Input to Output
4000
1500
1500
VAC
VAC
VAC
Input to PE
Isolation
Output to PE [Standard version only (RTN≠GND)]
PROTECTION:
Characteristic
Typ. Max.
Condition
Hiccup Mode, (Non-Latching, Auto-Recovery)
Hiccup Mode, (Non-Latching, Auto-Recovery)
Latch Mode
Min.
Unit
Short Circuit
Over Temperature
Overvoltage
105
110
150
154
%
%
Overload / Overcurrent
Hiccup Mode, (Non-Latching, Auto-Recovery)
ENVIRONMENTAL:
Characteristic
Typ. Max.
Condition
Derate Linearly From 100% Load at 40°C to 50% Load at 70°C
Surrounding Air Temperature
All Condition
10-95% RH (Non-Condensing)
Min.
-40
-40
Unit
°C
°C
Operating Temperature
Storage Temperature
Temperature Coefficient
Operating Humidity
Storage Humidity
70
85
±0.04
95% RH
95% RH
m
%/°C
10
0
Up to 5,000 meters
Non-Operating, 5-500 Hz, 2.09 Grms, 20 minute for Each Three Axis
Operating Altitude
Vibration
5000
2.09
G
1
2024.06