H HiTRON
Universal input AC-DC Medical and ITE application open frame
internal switching power supplies convection-cooled
200-250 Watts dual output HICM250 series
Features
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
6x4 inch Compact size
Very High Efficiency up to 90%
200-250W Convection/300W Forced-cooled
U-Bracket or Box format optional
Medical and ITE application
Class I construction
Specification
Input
General
Input Voltage
Input Frequency
Input Current
90-264VAC
Efficiency
Typical 90-92% (depending on model)
90-110KHz (depending on model)
IEC60601-1 and IEC60950-1
LLC circuit
47-63Hz
Switching Frequency
Dielectric Withstand
Circuit Topology
Transient Response
Typical 2-2.45A at 115VAC
Typical 1-1.2A at 230VAC
Typical 18.6A rms at 230VAC
Typical 0.97 at 230VAC
B-S connector
Inrush Current
Power Factor
Output voltage returns in less than
0.01-0.2mS for a 25% load change
Available
(depending on model)
Input Connector
Power OK
Earth Leakage Current Less than 0.15mA
Remote ON/OFF
Power Density
Construction
Available
5.3-6.6W / Cubic Inch
Output
Output Connector
Line Regulation
Load Regulation
Total Regulation
Noise & Ripple
Adjustability
B-S connector
U-Bracket and Box format optional
Typical 0.1%
Environmental
Operating Temperature
(Refer to the derating chart)
Storage Temperature
Cooling
Typical ±1%
-25°C to +70°C derate with
derating
Typical ±2%
Typical 1% peak to peak
Available at V1
27-30mS min. at 230VAC
-30°C to +85°C
Convection-cooled: 200-250W
Forced-cooled: 300W with 15CFM
10-95% RH, non-condensing
5-95% RH
Hold-up Time
Protection
Over Voltage
Over Load
Operating Humidity
Storage Humidity
Safety/EMC
Emissions (conducted)
Harmonic Current
Safety Standard
Built-in (Latch)
Typical set about 140-175% of
rating output wattage
Installed by NTC
CISPR EN55011 & EN55032 Class B
IEC61000-3-2
(depending on model)
Over Temperature
IEC60601-1/IEC60950-1Class I
Notes:
(1) All measurements are at nominal input, full load, and +25°C unless otherwise specified.
(2) Load regulation is measured at 115VAC or 230VAC in percentage to indicate the change in output voltage as the load varied from half load to full load (±%).
(3) The power supply is considered a component installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives.
(4) Due to requests in market and advances in technology, specifications subject to change without notice.
Version
0 Edited on April 14 2017
-1-