H HiTRON
Universal input harmonic correction (PFC) AC-DC medical & ITE application
external desktop switching adapter 150 Watts green power
single output HEMP156G series
Features
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
ꢀ
Energy Efficiency Level VI
Medical and ITE application
Wide operating temperature range
Up to 94% high efficiency
Class I construction
CE marking compliance
Specification
Input
Protection
Input Voltage
Input Frequency
Input Current
90-264VAC
Over Load
Typical set at about 130% of
rating output wattage
NTC
47-63Hz
Typical 1.6A at 115VAC
Typical 0.8A at 230VAC
27-30Arms at 230VAC 50Hz
Typical 0.95 at full load
3 pole IEC320-C14(DT7)
3 pole IEC320-C6(DT7L)
Less than 0.2mA
Over Temperature
General
Efficiency
Inrush Current
Power Factor
Typical 91-94% (depending on model)
85KHz
Switching Frequency
Dielectric Withstand
Transient Response
Input Connector
IEC60601-1 and IEC60950-1
Output voltage returns in less than
1mS for a 25% load change
4.37W / Cubic inch
Earth Leakage Current
Enclosure Leakage Circuit Less than 0.02mA
Power Density
No Load Power
Less than 0.21W
Construction
Desktop format
Output
Environmental
Operating Temperature
Storage Temperature
Cooling
Output Connector/Plug
Line Regulation
Load Regulation
Total Regulation
Noise & Ripple
Adjustability
Optional
-25°C to +60°C (Refer to derating chart)
-40°C to +85°C
Typical 0.1%
Typical ±2-3%
Convection-cooled
5000m
Typical ±3-4%
Operating Altitude
Operating Humidity
Storage Humidity
Typical 1% peak to peak
Not available
10-95% RH, non-condensing
5-95% RH
Hold-up Time
Typical 20mS at 115/230VAC
Safety/EMC
Emissions
EN55011 and EN55022 FCC Class B
IEC61000-3-2
Protection
Over Voltage
Built-in (Latch)
Installed
Harmonic Current
Safety Standard
Over Current
IEC60601-1 & IEC60950-1 Class I
Notes:
(1) All measurements are at nominal input, full load, and +25°C unless otherwise specified.
(2) Load regulation is measured at 115VAC or 230VAC in percentage to indicate the change in output voltage as the load varied from half load to full load ( %).
(3) The exact obtainable load regulation depends upon the output cord selected and load current.
(4) Due to requests in market and advances in technology, specifications subject to change without notice
Version
0 Edited on Aug 1, 2016
-1-