GSM28 Medical
28 Watt Global Performance Switchers
GLOBAL PERFORMANCE SWITCHERS
FEATURES:
D C P o w e r S u p p l i e s
• Industry’s smallest 28 W medically approved switcher
• Compact size (4.00” x 2.59” x 0.92”)
• Wide-range ac input: 85-264 Vac
• Less than 25 μA leakage current @ 120 Vac
• Approved to UL2601-1, EN60601-1
• EMI to FCC, CISPR 11 Class B/IEC601-1-2
• Overvoltage protection standard
• RoHS Compliant Models Available (G suffix)
•
marked to LVD
SPECIFICATIONS
Temperature Coefficient
0.03%/°C typical on all outputs.
Ac Input
85-264 Vac, 47-63 Hz single phase.
Environmental
Input Current
Designed for 0 to 50°C operation at full rated output power; derate output current
and total output power by 2.5% per °C above 50°C.
Maximum input current at 120 Vac, 60 Hz with full rated output load is 0.85 A.
Hold-up Time
Medical EMI/EMC Compliance
15 ms minimum from loss of ac input at full load, nominal line (120 Vac).
All models include built-in EMI filtering to meet the following emissions require-
ments:
Output Power
Normal continuous output power is 28 W, 32 W peak for 60 s maximum dura-
tion, 10% duty cycle. Factory set to begin power limiting at approximately 35
W.
EMI SPECIFICATIONS
COMPLIANCE LEVEL
Conducted Emissions
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Line Frequency Harmonics
EN55011 Class B; FCC Class B
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-3, 3 V/meter
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV diff., 2 kV com.
EN61000-3-2 Class A
Overload Protection
Fully protected against short circuit and output overload. Short circuit pro-
tection is cycling type power limit.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with
scope probe directly across output terminals of the power supply with load
Earth Leakage Current
Leakage current measured in the Gnd wire connection when measured per
EN60601-1 or UL2601-1 is as follows:
terminated with 0.1 μF capacitor.
Overvoltage Protection
Built in with firing point set per ratings table. OVP firing reduces voltage to
less than 50% of nominal voltage in 50 ms.
Medical
Model
Normal
SingleFault
Leakage
Test
Test
Leakage
Voltage
Method
Voltage Adjustment
Factory set with fixed resistors to maximize reliability.
GSM28
GSM28
25μA
50 μA
45 μA
90 μA
132 Vca/60 Hz
264 Vca/50 Hz
UL2601-1
IEC60601-1
Efficiency
70% minimum for the 5.1 V model at full rated load, nominal input voltage.
Efficiency increases as output voltage increases.
Medical Medical Safety
SL Power Electronics Corp. declares under our sole responsibility that all GSM
models are in conformity with the applicable requirements of UL2601-1 Patient
Care Equipment, CSA-C22.2 No. 234 (with additional tests to C22.2 No. 601.1 per
T.I.L. CA-08), EN60601-1.
Input Protection
Internal ac fuse provided on all units. Designed to open only if a catastrophic
failure occurs in the unit.
Inrush Current
Inrush limited by internal thermistor. The inrush at 230 Vac, averaged over the
first ac half-cycle under cold start conditions will not exceed 32 A.
Minimum Load Requirement
5% of full load rating
Transient Response 3.5% max. dev. 50% load step at 0.2 A/μs. Recovery to
0.5% within 500 μs.
All specifications are typical at nominal input, full load at 25OC unless otherwise stated