GPC130 Commercial/GPM130 Medical
130 Watt Global Performance Switchers
FEATURES:
• Wide-range ac input 85-264 Vac
• 2-year warranty
• Commercial Approved to UL1950, IEC950,
EN60950 and CSA22.2-234 L3
• Medical Approved to UL2601-1, IEC601-1 and
CSA22.2 No. 601
• Conducted EMI exceeds FCC Class B and CISPR 22
Class B (Commercial models) and CISPR 11 Class B
(Medical models)
•
marked to LVD
Power Fail
SPECIFICATIONS:
TTL- or CMOS-compatible output goes low (< 0.5V) 5 ms
before output voltage drops more than 4% below nominal
voltage upon loss of ac power.The signal is factory set to
trip on 84 to 94Vac brown-out depending upon incoming line
impedance and distortion. Other settings are available to the
user through adjustment of built-in potentiometer.
Ac Input
85-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at 120Vac, 60 Hz with full rated output
load: 3.5 A
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode.
Measured with noise probe directly across output terminals of
the power supply.
Hold-UpTime
20 ms minimum from loss of ac input at full load, nominal line
(115Vac).
Transient Response
Output Power
100 W continuous, 165W peak;130W with 26 cfm air flow.Peak
ratings are for 60 s maximum duration, 10% duty cycle. During
peak load condition, output regulation may exceed total
regulation limits.
Main output—500µSec typical response time for return to within
0.5% of final value for a 50% load step change. ∆i/∆t<0.2 A/µs.
Maximum voltage deviation is 3.5%. Startup/shutdown overshoot
less than 3%.
Overload Protection
Remote Sense
Provided as a standard feature on output 1.
Fully protected against short circuit and output overload. Short
circuit protection is cycling type power limit on outputs 1 & 2;
foldback type on outputs 3 & 4. Recovery after fault is automatic.
See output ratings chart for additional notes or conditions.
Voltage Adjustment
Built-in potentiometer adjusts voltage ±5% on outputs 1 & 2.
Overvoltage Protection
Main outputs: 124% ± 12% typical.
Overload Protection
Factory set to begin power limiting at approximately 170W.
Efficiency
EMI/EMC Compliance
70% at full rated load, nominal input voltage, depending on model
and load distribution.
All models include built-in EMI filtering to meet the following
emissions requirements:
EMI SPECIFICATIONS
COMPLIANCE LEVEL
Turn-onTime
Conducted Emissions GPC130 EN55022 Class B; FCC Class B
Conducted Emissions GPM130 EN55022 Class B; FCC Class B
Less than 1 second at 120Vac, 25°C (inversely proportional to
input voltage and thermistor temperature).
Static Discharge
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-3, 3 V/meter
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Input Protection
Internal ac fuse provided. Designed to blow only if a catastrophic
failure occurs in the unit.
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV diff., 2 kV com.
Commercial Leakage Current
0.7 mA 254 Vac @ 60 Hz input.
Inrush Current
Inrush is limited by internal thermistors.Inrush at 240Vac under
cold start conditions will not exceed 34 A.
Commercial Safety
Approved to UL1950, CSA22.2 No. 234 Level 3, IEC950 and
EN60950. UL file #E135803 commercial; CSA #LR46516 all
models.The output(s) are intended for safety earthed Signal
Output and Intermediate Circuits only. All dc outputs are SELV
under normal and single fault conditions.
Temperature Coefficient
0.03%/°C typical on all outputs.
Thermal Shutdown
Provided as a standard feature. Designed to protect unit from
prolonged overtemperature.
Medical Leakage Current
35 µA 254 Vac @ 60 Hz input.
Environmental
Medical Safety
Designed for 0 to 50°C operation at full rated output power;
derate output current and total output power by 2.5% per °C
above 50°C. See Environmental and Packaging Specifications
on next page.
Approved to UL2601, CSA22.2 No. 601 Level 3 and IEC601. UL
file E116994;CSA #LR46516.The output(s) are intended for
safety earthed Signal Output and Intermediate Circuits only.The
output(s) are not acceptable for patient connection without
additional isolation.All dc outputs are SELV under normal and
single fault conditions.
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 • 805-486-4565 • FAX 805-487-8911 • www.condorpower.com