40W Single Output
Medical & Industrial Grade
GB40 Family
FEATURES AND BENEFITS
Medical
Approved to CSA/EN/IEC/UL60601-1, 3rd Edition
40W Open Frame and PCB-mount Power Supply
2.0” x 4.0” x 1.0” Package
E-cap Life of >8 Years
>1,000,000 Hours MTBF
3 Year Warranty
Universal Input 90-264VAC
<0.1W No Load Input Power
Approved to CSA/EN/IEC/UL62368-1
Test &
Measurement
Meets Class B Radiated & Conducted EMI, with
Margin
Industrial
Meets Heavy Industrial and IEC60601-1-2 4th
Edition Levels of EMC
Note: *Consult Factory for compliance information.
MODEL SELECTION
Output
Power
Rated
Current
Ripple &
Noise1
Line
Regulation
Load
Regulation
Input
Class/Termination
Output
Termination
Model Number2
Volts
Class I (Grounded) input,
3-pin AMP/Molex type
connector
Change “K” to “C” for
class II input
Change “K” to “P” for PCB
mount pins, class I input
GB40S05K01
GB40S09K01
GB40S12K01
GB40S18K01
GB40S24K01
GB40S48K01
75mV pk-pk
90mV pk-pk
120mV pk-pk
120mV pk-pk
240mV pk-pk
480mV pk-pk
5.0V
9.0V
5.0A
4.0A
25W
±1%
±1%
±1%
±1%
±5%
±5%
±5%
±5%
4-pin AMP/Molex type
connector for “K” and “C”
versions
36W
3.4A
2.22A
1.7A
40W
40W
12.0V
18.0V
PCB mount pins for
“P” and “V” versions
24.0V
48.0V
40W
40W
±1%
±1%
±5%
±5%
Change “K” to “PCB mount
pins”, class II input
0.83A
Note: 1. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uF & 47uF parallel capacitor.
2. Other output voltages available, consult factory.
3. All specifications are typical at 230VAC, full load, at 25°C ambient unless noted.
INPUT
OUTPUT
Input Voltage and
Turn On Time
Hold-Up Time
<700ms
100-240VAC, ±10%, 47-63Hz, 1Ø
Frequency
20ms/100VAC at full load
Input Current
Inrush Current
Input Fuses
115VAC: 1.2A, 230VAC: 0.6A
25W-40W continuous – See models chart for specific
voltage model ratings
Output Power
Output Voltage
264VAC, cold start: will not exceed 40A peak
3.15A, 250VAC fuse in both line and neutral
See models chart
Earth Leakage Current
(Input to Ground)
<500μA@264VAC, 60Hz, NC
<1mA@264VAC, 60Hz, SFC
500μs resp. time for return to w/in 0.5% of final value
for any 50% load step from 5% to 100% of rated load,
∆i/∆t<0.2A/μs
Transient Response
Earth Leakage Current
(Output to Ground)
<100μA@264VAC, 60Hz, NC
<500μA@264VAC, 60Hz, SFC
Max voltage deviation is +/-3.5%
Regulation
See models chart
Efficiency
>88%, typical
Note: All specifications are typical at 230VAC input, full load, at 25°C ambient unless noted.
0.9, min., 230VAC, 80-100% load vector, 25°C
ambient
Power Factor
Note: All specifications are typical at 230VAC input, full load, at 25°C ambient unless noted.
GB40 Datasheet v0818
Copyright © 2018 SL Power Electronics Corp. - All rights reserved.
Page 1