IEC Inlet Filters FN 280
DATASHEET
Compact filtered power entry module
Complies with IEC/EN 60601-1
Rated currents up to 10 A
Single or dual-fuse holder
Fuses Ø5 x 20 mm
Approvals
2-pole rocker switch
General purpose application
Optional earth line choke (E type)
Optional medical versions (B type)
(CQC except HI-types)
The FN 280 power entry module combines an IEC
inlet, a mains filter with single or dual- fuse holder and
a 2-pole rocker switch in a small form factor.
Choosing FN 280 product line brings you the rapid
availability of a standard filter associated with the
necessary safety acceptances. Standard IEC
connector filters are a practical solution helping you
to pass EMI system approval in a short time. A wide
selection on amperage ratings, fuse options,
mounting possibilities and filters for medical
applications are designed to offer you the desired
solution.
Performance indicators
Features and benefits
High conducted attenuation performance, based
on chokes with high saturation resistance and
excellent thermal behavior
Technical specifications
Rear/front or snap-in mounting
Single or dual-fuse holder
250 V, 50/60 Hz
DC to 400 Hz
Maximum continuous operating voltage
Operating frequency
2-pole rocker switch
Custom-specific versions are available on request
1 to 10 A @ 40°C max.
Rated currents
P –> PE 2000 VAC for 2 sec (standard types)
P –> PE 2500 VAC for 2 sec (B types)
P –> N 760 VAC for 2 sec
High potential test voltage
Typical applications
Portable electrical and electronic equipment
Consumer goods
IP 40 according to IEC 60529
-25°C to +85°C (25/85/21)
Protection category
Temperature range (operation and storage)
Design corresponding to
Flammability corresponding to
MTBF @ 40°C/230 V (Mil-HB-217F)
Fuse holder
UL 1283, CSA 22.2 No. 8 1986, IEC/EN 60939
UL 94 V-2 or better
EDP and office equipment
Single-phase power supplies, switch-mode power
supplies
1,500,000 hours
1 or 2 fuses (Ø5 x 20 mm) (certified to IEC 60127-6)
Test and measurement equipment
Medical equipment
Power acceptance @ amb. temperature
Marking
max. 250 V ~ (text is print. below fuse symbol)
Rocker switch description
Function
Typical electrical schematic
2-pole, dark not illuminated
Marking I - 0
Inrush current 82 A
Electrical specifications
6,000 on-off operations according to UL 1054, TV5
10,000 on-off operations according to ENEC
Switch ratings
10 A, 125 VAC; 10 A, 250 VAC; 1/3 HP
10 A (4 A), 250 VAC*
USA (UL) and Canada (C-UL)
Europe (ENEC)
50,000 cycles
Mechanical life
* Value in () relates to the inductive current charge: cosφ = 0.65