IN-SITU  
DATA SHEET  
ZIRCONIA OXYGEN ANALYZER  
FOR HAZARDOUS LOCATION  
ZFKE, ZKME  
This oxygen analyzer is used to continuously measure  
oxygen concentration in noncombustible exhaust gas of  
industrial boilers or furnaces, and is ideally suited for com-  
bustion management and control.  
The analyzer system is comprised of the detector and  
converter coupled together as a complete system. Detec-  
tor setting configuration includes the flow guide tube and  
detector sensor. The flow guide tube is inserted directly  
into the gas and directs gas to the sensor for measurement.  
The converter (ZKME) is comprised of the signal proces-  
sor, input/output and communications, display and system  
controls.  
Detector (ZFKE)  
The converter is equipped with advanced functional-  
ity such as performing the sensor diagnostics and sensor  
recovery function, so the detector can be used within long  
term stability.  
FEATURES  
Converter (ZKME)  
1. Gas sampling device is unnecessary  
For quick response, insert the detector directly into the  
staek. Gas sampling functions such as a gas aspirator  
and a dehumidifier are not required.  
SPECIFICATIONS  
General Specifications  
2. Easy maintenance  
Measuring object: Oxygen in noncombustible gas  
Measuring method:  
Directly insert type zirconia system  
Measuring range: 0 to 2 ··· 0 to 50 vol% O2  
2 ranges available in 1 vol% O2 steps  
The sensor equipped with the detector, has unit construc-  
tion, it is easy to replace.  
By separating the detector and the flow guide tube, filter  
replacement is easy.  
3. More reliable than sensor diagnosis, sensor recoverable  
function  
Repeatability:  
Linearity:  
Within ±0.5%FS  
Within ±2%FS  
Depending on the components in the measurement  
gas, the characteristics of the sensor might deteriorate.  
The equipment includes sensor recovery function elec-  
tronically, checking the deterioration status of the sensor  
depletion.  
Response time:  
Within 4 to 7 sec, for 90% (from calibra-  
tion gas inlet)  
Warmup time:  
Analog output:  
approx. 10 min  
4 to 20mA DC (allowable load resis-  
tance less than 500Ω) or 0 to 1V DC  
(output resistance more than 100Ω)  
Rated voltage;  
Therefore, it has high reliability and long-lasting stability.  
4. Safe and secure  
Power supply:  
System detects thermocouple break for heater control on  
the sensor side. Safety functions of isolating power sup-  
ply to the detector or isoralting power via external contact  
input are also.  
100 to 120V AC (operating voltage 90  
to 132V AC)  
200 to 240V AC (operating voltage  
190 to 264V AC)  
5. Easy operation  
Rated frequency; 50/60Hz  
Power consumption:  
Maximum 240VA (Detector: approx.  
The operation and setting for the converter can be per-  
formed interactively, and available as English, Japanese  
or Chinese for language display.  
200VA, Converter: approx. 40VA)  
Normal 70VA (Detector: approx. 50VA,  
Converter: approx. 20VA  
EDS3-139b  
Date  
Apr. 1, 2011  

与EDS3-139B相关器件

型号 品牌 描述 获取价格 数据表
EDS3-141A FUJI NDIR TYPE INFRARED GAS ANALYZER(3-COMPONENT ANALYZER)

获取价格

EDS3-143 FUJI INFRARED GAS ANALYZER

获取价格

EDS3-144 FUJI INFRARED GAS ANALYZER(HIGH PERFORMANCE MODEL)

获取价格

EDS3-145 FUJI INFRARED GAS ANALYZER(LOW-CONCENTRATION MEASUREMENT TYPE)

获取价格

EDS-316 MOXA 16-port unmanaged Ethernet switches

获取价格

EDS-316-MM-SC MOXA 16-port unmanaged Ethernet switches

获取价格