DTGMPU70 SERIES
70 Watts
AC/DC Medical Desktop Power Supply
Single Output
Rev C
FEATURES
• Universal Input Voltage Range • Over Voltage, Over Current, and Short Circuit
of 100~240VAC
Protection
• Low Leakage Current
• Means of Patient Protection Design
• Meets EISA 2007/DoE (VI) & • Meets 60601-1 3rd Edition
EU ErP/CoC (5)
• C14 or C6 AC Inlet
• UL: ES60601-1, CSA:C22.2 NO.60601-1, CB: IEC
60601-1, and EN: EN60601-1 Safety Approvals
DESCRIPTION
The DTGMPU70 series of AC/DC medical desktop power supplies offers 70 watts of
output power in a 5.75” x 2.97” x 1.69” package. This series consists of single output
models with a universal input voltage range of 100~240VAC and either a C14 or C6
AC inlet. Each model in this series has low leakage current, meets EISA 2007/DoE
(VI) & EU ErP/CoC (5), and has over voltage, over current and short circuit protection.
This series has UL: ES60601-1, CSA:C22.2 NO. 60601-1, CB: IEC 60601-1, and EN:
EN60601-1 safety approvals.
Size: 5.75in x 2.97in x 1.69in (146mm x 75.5mm x 43mm)
MODEL SELECTION TABLE
Output Current
Min Load Max Load
Efficiency(2)
Input Voltage
Range
Output
Voltage
Max. Output
Power
No Load
Power Consumption at Output
Measured
Model Number(1)
Ripple Max
DoE (VI) CoC (5)
DTGMPU70X-3
DTGMPU70X-4
DTGMPU70X-5
DTGMPU70X-6
DTGMPU70X-7
DTGMPU70X-8
11~13VDC
13~16VDC
16~21VDC
21~27VDC
27~33VDC
33~48VDC
5.39A
4.38A
3.34A
2.60A
2.13A
1.22A
6.37A
5.39A
4.38A
3.34A
2.60A
2.13A
70W
70W
70W
70W
70W
70W
120mV
120mV
180mV
180mV
240mV
240mV
88%
88%
88%
88%
88%
88%
89%
89%
89%
89%
89%
89%
12
15
18
100~240VAC
<0.15W
24
32
48
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
INPUT SPECIFICATIONS
Input Voltage Range
Input Frequency
Input Current
TEST CONDITIONS
Min
Typ
100
±5
Max
Unit
100
50
1.8
240
60
0.8
VAC
Hz
A
A
mA
Inrush Current
@230VAC, 25ºC, Cold Start
@240VAC/50Hz
Leakage Current
OUTPUT SPECIFICATIONS
Output Voltage
Line Regulation
Load Regulation
Output Power
Output Current
No Load Power Consumption
Ripple
0.1
See Table
±1
For any input voltage change between input voltage range
Variations from minimum to maximum output current
%
%
See Table
See Table
0.15
%
See Table
Maximum excursion of 4% or better on all models.
Recovering to 1% of final value within 500uS after a 25% step load change.
Transient Response
Hold-Up Time
8
mSec
PROTECTION
Short Circuit Protection
Hiccup Mode
Hiccup Mode
Rated Output Current
Protected by Zener Diode
Rated Output Voltage
Automatic Recovery
Automatic Recovery
Over Current Protection
Over Voltage Protection
110
110
%
%
140
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
-20
-40
5
+40
+85
95
ºC
ºC
%
Relative Humidity
Derating
MTBF
Non-Condensing
Derated from 100% at +40ºC linearly to 50% at 70ºC
@Full Load, 25ºC ambient
100,000
Hours
Wall Industries, Inc. • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
website: www.wallindustries.com • e-mail: sales@wallindustries.com
10/1/2019
Page 1 of 3