DTGMPU18 SERIES
18 Watts
AC/DC Medical Desktop Power Supplies
Single Output
Rev B
FEATURES
• Universal Input Voltage Range of • Over Voltage, Over Current, and Short Circuit
100~240VAC
• Single Outputs
• Possess Risk Analysis Report
• Means of Patient Protection
• IEC-320-C14, IEC-320-C18, or
IEC-320-C16
Protection
IEC-320-C14
• For use in Personal Hygiene and Health Care
Appliances
• UL: ES60601-1, CSA: C22/2 NO. 60601-1,
CB: IEC60601-1, and EN: EN60601-1 Safety
Approvals
DESCRIPTION
The DTGMPU18 series of AC/DC medical desktop power supplies offers 18 watts of
output power in a 3.94” x 2.30” x 1.29” package. This series consists of single output
models with universal input voltage range of 100~240VAC and different options
available. Either an IEC-320-C14, IEC-320-C18, or IEC-320-C6 AC inlet and there
are several optional output connectors available. Each model in this series has over
voltage, over current, and short circuit protection while 5~6 and 8~58VDC models
meet EISA 2007/DoE (VI) & EU ErP/CoC (5).
IEC-320-C18
IEC-320-C6
Size: 3.94in x 2.30in x 1.29in (100mm x 58.5mm x 32.8mm)
MODEL SELECTION TABLE
Output
Input
Max.
Output
Power
Efficiency
Efficiency
Level
No Load
Power
Output
Current
Ripple
Max Regulation
Load
Measured
at Output
Model Number
Voltage
Range
Voltage
Min
Max
DoE (VI) CoC (5) (DoE/CoC) Consumption
Load Load
2.50A 3.00A
1.87A 2.50A
DTGMPU18X-1
DTGMPU18X-1-1
DTGMPU18X-2
DTGMPU18X-3
DTGMPU18X-4
DTGMPU18X-5
DTGMPU18X-6
DTGMPU18X-7
DTGMPU18X-8
5~6VDC
6~8VDC
8~11VDC
15W
15W
80mV
80mV
±5%
±5%
±5%
±5%
±5%
±5%
±5%
±3%
±3%
>81.39% >81.84%
VI, 5
V, 4
<0.075W
<0.3W
5
7.5
9
12
15
18
24
28
48
>79.2%
>85%
>85%
>85%
>85%
>85%
>85%
>85%
>79.2%
>85.45%
>85.45%
>85.45%
>85.45%
>85.45%
>85.45%
>85.45%
1.64A 2.25A 18.04W 100mV
VI, 5
VI, 5
VI, 5
VI, 5
VI, 5
VI, 5
VI, 5
<0.075W
<0.075W
<0.075W
<0.075W
<0.075W
<0.075W
<0.075W
11~13VDC 1.38A 1.64A 18.04W 120mV
13~16VDC 1.13A 1.38A 18.08W 150mV
16~21VDC 0.86A 1.13A 18.08W 150mV
21~27VDC 0.67A 0.86A 18.09W 150mV
27~33VDC 0.55A 0.67A 18.15W 240mV
33~58VDC 0.32A 0.55A 18.56W 240mV
100~240
VAC
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
INPUT SPECIFICATIONS
Input Voltage Range
Input Frequency
TEST CONDITIONS
Min
Typ
Max
Unit
100
47
0.45
240
63
0.2
VAC
Hz
A
Input Current
@115VAC, 25ºC, Cold Start
@230VAC, 25ºC, Cold Start
@240VAC/50Hz
50
90
Inrush Current
A
Leakage Current
OUTPUT SPECIFICATIONS
Output Voltage
Line Regulation
Load Regulation
Output Power
0.1
mA
See Table
±1
See Table
See Table
See Table
See Table
For any input voltage change between input voltage range
Variations from minimum to maximum output current
%
Output Current
Ripple
Maximum Excursion of 45 or better on all models.
Recovering to 1% of final value within 500uS after a 25% step load change.
Transient Response
Set Up Time
Hold Up Time
Rise Time
Temperature Coefficient
PROTECTION
@Full Load
@Full Load
@Full Load
All Outputs
1000
10
50
mS
mS
mS
±0.04
%/ºC
Short Circuit Protection
Hiccup Mode
Hiccup Mode
Rated Output Current
Protected by Zener diode
Rated Output Voltage
Automatic Recovery
Automatic Recovery
Over Current Protection
Over Voltage Protection
110
110
%
%
140
Wall Industries, Inc. • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
website: www.wallindustries.com • e-mail: sales@wallindustries.com
10/1/2019
Page 1 of 3