DTAM90/100 SERIES
90~110 Watts
Medical AC/DC Desktop Power Supplies
Single Outputs
Rev G
FEATURES
• Class I
• 3-Pole IEC-320-C14 AC Inlet
• RoHS Compliant
• 90~264VAC Input Voltage Range
• Up to 110 Watts Output Power
• Single Outputs
• No Load Power Consumption <500mW
• Power Factor: >0.95 at 115VAC and
Full Load
• Earth Leakage Current: <266µA at
264VAC
• Power ON LED Indicator
• Over Voltage, Over Load, and Short
Circuit Protection
• Meets CEC, ERP, EISA, and Energy Star
Level V Requirements
• UL60601-1: 3rd Edition Medical Approvals
• Optional Output Connectors Available
DESCRIPTION
The DTAM90/110 series of Class I desktop AC/DC power supplies delivers up to 110
Watts of output power in a 6.69” x 2.55” x 1.52” package. This series consists of
single output models with an input voltage range of 90-264VAC and a 3-pole IEC-320-
C14 AC inlet. These supplies are protected against over voltage, over load, and short
circuit conditions. The DTAM90/110 series is RoHS compliant and meets CEC, ERP,
EISA, and Energy Star Level V requirements. These units also have UL60601-1: 3rd
edition medical approvals.
Size: 6.69in x 2.55in x 1.52in (170mm x 64.8mm x 38.5mm)
MODEL SELECTION TABLE
Model Number
Input Voltage Range
90~264VAC
Output Voltage
Output Current Ripple & Noise(1) Output Regulation Output Power Avg. Efficiency
DTAM-090A-1Y120G
DTAM-090A-1Y135G
DTAM-100A-1Y150G
DTAM-110A-1Y190G
DTAM-110A-1Y200G
DTAM-110A-1Y240G
12VDC
13.5VDC
15VDC
19VDC
20VDC
24VDC
7.5A
6.67A
6.67A
5.8A
120mVp-p
135mVp-p
150mVp-p
190mVp-p
200mVp-p
240mVp-p
±5%
±5%
±5%
±3%
±3%
±3%
90W
90W
100W
110W
110W
110W
>87%
5.5A
4.6A
SPECIFICATIONS
All specifications are based on 25°C, Nominal Input Voltage, and Maximum Output Current unless otherwise noted.
We reserve the right to change specifications based on technological advances.
SPECIFICATION
TEST CONDITIONS
Min
Typ
Max
Unit
INPUT SPECIFICATIONS
Input Voltage Range
Input Frequency
90
47
115/230
50/60
264
63
2
VAC
Hz
@90VAC
@264VAC
Input Current
Arms
1
@115VAC, cold start, and 25ºC
@230VAC, cold start, and 25ºC
50
100
500
Inrush Current
No Load Power Consumption
Power Factor
mW
-
@115VAC and Full Load
@230VAC and Full Load
0.95
0.90
OUTPUT SPECIFICATIONS
Output Voltage
See Table
Output Regulation
Output Power
Output Current
See Table
See Table
See Table
Minimum Load
0
mA
ms
Measured at 20MHz limited bandwidth with 0.1µF ceramic and 10µF
electrolytic capacitors across the output
@Full Load, 115VAC
Ripple & Noise (20MHz bandwidth)
See Table
Hold-Up Time
>10
PROTECTION
Short Circuit Protection
Over Load Protection
Over Voltage Protection
ENVIRONMENTAL SPECIFICATIONS
Operating Case Temperature
Storage Temperature
Humidity
Auto-Recovery
Auto-Recovery
Shutdown and Latch Off, AC Recycle
Derate linearly 2.5% per ºC from 41 to 60ºC
@Full Load and 25ºC ambient temperature
0
-10
10
60
+70
95
ºC
ºC
% RH
hours
MTBF
100,000
Wall Industries, Inc. • Tel: 603-778-2300 • Toll Free: 888-597-9255 •
website: www.wallindustries.com • e-mail: sales@wallindustries.com
6/25/2019
Page 1 of 4