ADS1192IRSMR PDF预览

ADS1192IRSMR

更新时间: 2024-11-05 12:36:11
品牌 Logo 应用领域
德州仪器 - TI /
页数 文件大小 规格书
67页 1853K
描述
Low-Power, 2-Channel, 16-Bit Analog Front-End for Biopotential Measurements

ADS1192IRSMR 技术参数

是否无铅: 不含铅是否Rohs认证: 符合
生命周期:Active包装说明:HVQCCN,
Reach Compliance Code:compliantHTS代码:8542.39.00.01
Factory Lead Time:6 weeks风险等级:5.55
其他特性:IT ALSO REQUIRES 1.8 V模拟集成电路 - 其他类型:ANALOG CIRCUIT
JESD-30 代码:S-PQCC-N32JESD-609代码:e4
长度:4 mm湿度敏感等级:2
信道数量:2功能数量:1
端子数量:32最高工作温度:85 °C
最低工作温度:-40 °C封装主体材料:PLASTIC/EPOXY
封装代码:HVQCCN封装形状:SQUARE
封装形式:CHIP CARRIER, HEAT SINK/SLUG, VERY THIN PROFILE峰值回流温度(摄氏度):260
座面最大高度:1 mm最大供电电压 (Vsup):5.25 V
最小供电电压 (Vsup):2.7 V标称供电电压 (Vsup):3 V
表面贴装:YES温度等级:INDUSTRIAL
端子面层:Nickel/Palladium/Gold (Ni/Pd/Au)端子形式:NO LEAD
端子节距:0.4 mm端子位置:QUAD
处于峰值回流温度下的最长时间:NOT SPECIFIED宽度:4 mm
Base Number Matches:1

ADS1192IRSMR 数据手册

 浏览型号ADS1192IRSMR的Datasheet PDF文件第2页浏览型号ADS1192IRSMR的Datasheet PDF文件第3页浏览型号ADS1192IRSMR的Datasheet PDF文件第4页浏览型号ADS1192IRSMR的Datasheet PDF文件第5页浏览型号ADS1192IRSMR的Datasheet PDF文件第6页浏览型号ADS1192IRSMR的Datasheet PDF文件第7页 
ADS1191  
ADS1192  
www.ti.com  
SBAS566A DECEMBER 2011REVISED SEPTEMBER 2012  
Low-Power, 2-Channel, 16-Bit Analog Front-End for Biopotential Measurements  
Check for Samples: ADS1191, ADS1192  
The ADS1191/2 incorporate all of the features that  
1
FEATURES  
are commonly required in portable, low-power  
medical electrocardiogram (ECG), sports, and fitness  
applications.  
23  
Two Low-Noise PGAs and  
Two High-Resolution ADCs (ADS1192)  
Low Power: 335 μW/channel  
With its high levels of integration and exceptional  
performance, the ADS1191/2 family enables the  
creation of scalable medical instrumentation systems  
at significantly reduced size, power, and overall cost.  
Input-Referred Noise: 24 μVPP  
(150-Hz BW, G = 6)  
Input Bias Current: 1 nA  
Data Rate: 125 SPS to 8 kSPS  
CMRR: –95 dB  
The ADS1191/2 have a flexible input multiplexer per  
channel that can be independently connected to the  
internally-generated signals for test, temperature, and  
lead-off detection. Additionally, any configuration of  
input channels can be selected for derivation of the  
right leg drive (RLD) output signal. The ADS1191/2  
operate at data rates up to 8 kSPS. Lead-off  
detection can be implemented internal to the device,  
using the device internal excitation current  
sink/source.  
Programmable Gain: 1, 2, 3, 4, 6, 8, or 12  
Supplies: Unipolar or Bipolar  
Analog: 2.7 V to 5.25 V  
Digital: 1.7 V to 3.6 V  
Built-In Right Leg Drive Amplifier, Lead-Off  
Detection, Test Signals  
Built-In Oscillator and Reference  
The devices are packaged in a 5-mm × 5-mm, 32-pin  
thin quad flat pack (TQFP). Operating temperature is  
specified from –40°C to +85°C.  
Flexible Power-Down, Standby Mode  
SPI™-Compatible Serial Interface  
REF  
Operating Temperature Range: –40°C to +85°C  
Test Signals and  
Monitors  
Reference  
APPLICATIONS  
SPI  
Medical Instrumentation (ECG) including:  
Patient monitoring; Holter, event, stress,  
and vital signs including ECG, AED,  
telemedicine  
ADC1  
ADC2  
A1  
A2  
Oscillator  
MUX  
Control  
Sports and fitness (heart rate, respiration,  
and ECG)  
To Channel  
DESCRIPTION  
The ADS1191/2 are  
a family of multichannel,  
simultaneous sampling, 16-bit, delta-sigma (ΔΣ)  
analog-to-digital converters (ADCs) with a built-in  
programmable gain amplifier (PGA), internal  
reference, and an onboard oscillator.  
RLD  
1
Please be aware that an important notice concerning availability, standard warranty, and use in critical applications of  
Texas Instruments semiconductor products and disclaimers thereto appears at the end of this data sheet.  
2
3
SPI is a trademark of Motorola.  
All other trademarks are the property of their respective owners.  
PRODUCTION DATA information is current as of publication date.  
Products conform to specifications per the terms of the Texas  
Instruments standard warranty. Production processing does not  
necessarily include testing of all parameters.  
Copyright © 2011–2012, Texas Instruments Incorporated  

ADS1192IRSMR 替代型号

型号 品牌 替代类型 描述 数据表
ADS1192IRSMT TI

功能相似

Low-Power, 2-Channel, 16-Bit Analog Front-End for Biopotential Measurements

与ADS1192IRSMR相关器件

型号 品牌 获取价格 描述 数据表
ADS1192IRSMT TI

获取价格

Low-Power, 2-Channel, 16-Bit Analog Front-End for Biopotential Measurements
ADS1194 TI

获取价格

Low-Power, 8-Channel, 24-Bit Analog Front-End for Biopotential Measurements
ADS1194CPAG TI

获取价格

Eight Low-Noise PGAs and Eight High-Resolution ADCs (ADS1198)
ADS1194CPAGR TI

获取价格

暂无描述
ADS1194CZXGR TI

获取价格

Low-Power 8-Channel 16-Bit Analog Front-End for Biopotential Measurements
ADS1194CZXGT TI

获取价格

Low-Power 8-Channel 16-Bit Analog Front-End for Biopotential Measurements
ADS1194IPAG TI

获取价格

Low-Power 8-Channel 16-Bit Analog Front-End for Biopotential Measurements
ADS1194IPAGR TI

获取价格

Low-Power 8-Channel 16-Bit Analog Front-End for Biopotential Measurements
ADS1196 TI

获取价格

Low-Power, 8-Channel, 24-Bit Analog Front-End for Biopotential Measurements
ADS1196CPAG TI

获取价格

用于 ECG 的低功耗、6 通道、16 位模拟前端 | PAG | 64 | 0 to 7