Embedded Power for
Business-Critical Continuity
Rev. 03.19.08
AD50 Medical Series
1 of 2
AD50 Medical
Series
RoHS
50 Watts
Total Power:
50 Watts
Input Voltage: 90 - 264 VAC
# of Outputs:
single
Electrical Specifications
Input
Input range
Frequency
Inrush current
Input current
Efficiency
EMI/RFI
Safety ground
leakage current
90-264 Vac (wide range)
47-63 Hz
60A maximum @ 230VAC, cold start 25C°
2 A maximum
Special Features
• Universal AC input
• Fully regulated output
• EN61000-3-2 compliant
• Input power <74 watts
• Overcurrent and overvoltage
protection
80% typical
FCC Part 15, Class B & EN55022 (CISPR 22) Class B
300μA maximum@ 50/60 Hz, 264 VAC input - AD5012N3L
100μA maximum@ 50/60 Hz, 264 VAC input - AD5012N2L
• High efficiency
• High MTBF
Output
• IEC320 AC input receptacle
3 pin (type C6) - AD5012N3LM
2 pin (type C8) - AD5012N2LM
• Built in EMI filter
50 W
Maximum Power (Po)
Hold-up time
Overvoltage protection
17 ms. minimum at full load @ 100 Vac, 47 Hz
16 - 20 V
Latching type, recycle AC to reset.
Output short circuit protection auto recover
Overload protection @ 110 - 120% above maximum rating
• (CISPR 22 Class B)
• AC input fuse
• Complies with One Watt Input
Energy Star / Blue Angel
Requirement
Overcurrent protection
105° to 120°C Trip point; output will latch off
Thermal protection
Cable/connector
DC cable length 2.5 mm center plug
DC plug center +v
DC plug outer -v
• CE Mark EMC & LVD
Environmental Specifications
Operating temperature:
5° to +40°C ambient
Storage temperature:
Electromagnetic
susceptibility:
-40°C to +70°C
Designed to meet EN61000-4-2, -4, -5, level 4;
EN61000-4-3, -6, 10v/m; EN61000-4-8, -11; EN61000-3-3
and EN61000-3-2 Class A
Safety
CSAUL UL 60601-1
Humidity:
Operating; non-condensing 10% to 90% RH
CSA
VDE
60601-1
60601-1
MTBF demonstrated:
>300,000 hours at full load and 25°C ambient conditions
CE Mark LVD and EMC