P7600 Series TriMode Probes
®
TekConnect Interface - TekConnect scope/probe control and usability
Direct control from probe compensation box or from scope menu
Applications
Including, but not limited to:
MIPI M-PHY, PCI Express, Serial ATA, Serial Attached SCSI, Fibre
Channel, and HDMI
Combining TriMode probing with low noise
performance
TriMode probing enhances productivity by enabling differential, single-
ended, and common mode measurements with a single probe setup.
Connecting a probe to the device under test can be a time consuming
activity, especially if the probe has to be set up differently to make all the
necessary measurements. TriMode probing improves productivity by
reducing setup time, because only one setup is needed to make the three
different types of measurements. Switching between Differential Mode [A-
B], Single-Ended Mode [A-Gnd, B-Gnd], and Common Mode [((A+B)/2)-
Gnd] is as easy as a press of a button.
P7633 Low Noise TriMode Probe
The P7600 Series TriMode probes and the MSO/DPO70000DX and DPO/
DSA70000D Series Oscilloscopes are designed to deliver the industry's
lowest system noise levels. This high sensitivity is critical for being able to
make accurate measurements on low amplitude signals.
Key performance specs
Full 33 GHz bandwidth to the probe tip (P7633)
Industry's lowest probe and oscilloscope system noise
Key features
™
TriMode probe - one setup, three measurements without adjusting
probe tip connections
Differential
Single ended
Changing input modes on the probe is as easy as a press of a button
Common mode
As high speed serial data standards increase in speed and signal swings
shrink, probes with high bandwidth coupled with low noise and high
sensitivity are increasingly important for making accurate measurements.
Remote head design connectivity
Allows placement of probe amplifier close to the circuit being
measured
Coaxial adapters
2.92mm
SMP
P7500 probe tip adapter
Use P7500 Series TriMode solder-in probe tips
Probe automatically detects tips and applies DSP correction filters
Probe-specific S-parameters
12 Gb/s 200 mVpp signal measured with a P7633 and a P76CA-292C coaxial adapter
Unique DSP filters created from data stored in the probe
www.tek.com
1